Monitoring and correcting spatio-temporal variations of the MR scanner's static magnetic field

被引:85
作者
El-Sharkawy, Abdel Monem
Schar, Michael
Bottomley, Paul A.
Atalar, Ergin [1 ]
机构
[1] Bilkent Univ, Dept Elect & Elect Engn, TR-06533 Bilkent, Ankara, Turkey
[2] Johns Hopkins Univ, Dept Elect & Comp Engn, Baltimore, MD 21218 USA
[3] Johns Hopkins Univ, Sch Med, Russell H Morgan Dept Radiol & Radiol Sci, Baltimore, MD 21287 USA
[4] Philips Med Syst, Cleveland, OH USA
关键词
magnetic field; field homogeneity; magnet stability; field mapping; MR thermometry;
D O I
10.1007/s10334-006-0050-2
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
The homogeneity and stability of the static magnetic field are of paramount importance to the accuracy of MR procedures that are sensitive to phase errors and magnetic field inhomogeneity. It is shown that intense gradient utilization in clinical horizontal-bore superconducting MR scanners of three different vendors results in main magnetic fields that vary on a long time scale both spatially and temporally by amounts of order 0.8-2.5 ppm. The observed spatial changes have linear and quadratic variations that are strongest along the z direction. It is shown that the effect of such variations is of sufficient magnitude to completely obfuscate thermal phase shifts measured by proton-resonance frequency-shift MR thermometry and certainly affect accuracy. In addition, field variations cause signal loss and line-broadening in MR spectroscopy, as exemplified by a fourfold line-broadening of metabolites over the course of a 45 min human brain study. The field variations are consistent with resistive beating of the magnet structures. It is concluded that correction strategies are required to compensate for these spatial and temporal field drifts for phase-sensitive MR protocols. It is demonstrated that serial field mapping and phased difference imaging correction protocols can substantially compensate for the drift effects observed in the MR thermometry and spectroscopy experiments.
引用
收藏
页码:223 / 236
页数:14
相关论文
共 33 条
[1]   Variation correction algorithm: Analysis of phase suppression and thermal profile fidelity for proton resonance frequency magnetic resonance thermometry at 0.2 T [J].
Barkauskas, KJ ;
Lewin, JS ;
Duerk, JL .
JOURNAL OF MAGNETIC RESONANCE IMAGING, 2003, 17 (02) :227-240
[2]  
BEHINA B, 2002, CONC MAGN RES, V15, P101
[3]   SPATIAL LOCALIZATION IN NMR-SPECTROSCOPY INVIVO [J].
BOTTOMLEY, PA .
ANNALS OF THE NEW YORK ACADEMY OF SCIENCES, 1987, 508 :333-348
[4]  
CLARKE N, 2006, Patent No. 7015692
[5]  
de Zwart JA, 1999, MAGNET RESON MED, V42, P53, DOI 10.1002/(SICI)1522-2594(199907)42:1<53::AID-MRM9>3.0.CO
[6]  
2-S
[7]  
DEPOORTER J, 1995, MAGN RESON MED, V34, P359, DOI 10.1002/mrm.1910340313
[8]  
EMERIC PR, 2006, Patent No. 6992483
[9]   Magnetic field shift due to mechanical vibration in functional magnetic resonance imaging [J].
Foerster, BU ;
Tomasi, D ;
Caparelli, EC .
MAGNETIC RESONANCE IN MEDICINE, 2005, 54 (05) :1261-1267
[10]  
Froundlich D., 2003, US Patent, Patent No. [6559644 InSightec, 6559644, 6.559.644]