Influence of anode/filter material and tube potential on contrast, signal-to-noise ratio and average absorbed dose in mammography: a Monte Carlo study

被引:134
作者
Dance, DR
Klang, AT
Sandborg, M
Skinner, CL
Smith, IAC
Carlsson, GA
机构
[1] Royal Marsden NHS Trust, Joint Dept Phys, London SW3 6JJ, England
[2] Inst Canc Res, London SW3 6JJ, England
[3] Univ Gothenburg, Sahlgrens Univ Hosp, Dept Radiat Phys, SE-41345 Gothenburg, Sweden
[4] Linkoping Univ, Fac Hlth Sci, IMV, Dept Radiat Phys, SE-58185 Linkoping, Sweden
关键词
D O I
10.1259/bjr.73.874.11271898
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
The comparative performance of mammographic X-ray systems that use different anode/filter combinations has been assessed for screen-film and digital imaging. Monte Carlo techniques have been used to calculate average glandular dose as well as contrast and signal-to-noise ratio for imaging two test details. Five anode/filter combinations have been studied to establish the potential for dose saving or image quality improvement. For screen-film mammography, it was found that little benefit is gained by changing from a standard 28 kV molybdenum/molybdenum spectrum for breasts up to 6 cm thick. For thicker breasts, where the tube potential for the standard technique might be increased, 20% improvement in contrast can be achieved without dose penalty using molybdenum/rhodium or rhodium/rhodium spectra, whereas dose savings of more than 50% can be attained whilst maintaining contrast using tungsten/rhodium or rhodium/aluminium spectra. In digital mammography, a molybdenum/molybdenum spectrum delivers the lowest dose for a 2 cm breast, but gives the highest dose for thicker breasts. Tungsten/rhodium or rhodium/aluminium spectra provide the lowest doses at greater thicknesses. It is concluded that for screen-film mammography, molybdenum/molybdenum is the spectrum of choice for all but the thickest or most glandular breasts. In digital mammography, an alternative spectrum is preferable for breasts thicker than 2 cm.
引用
收藏
页码:1056 / 1067
页数:12
相关论文
共 27 条
[1]  
Berger M J, 1987, 873597 NBSIR US DEP
[2]   CALCULATION OF DOSE AND CONTRAST FOR 2 MAMMOGRAPHIC GRIDS [J].
DANCE, DR ;
PERSLIDEN, J ;
CARLSSON, GA .
PHYSICS IN MEDICINE AND BIOLOGY, 1992, 37 (01) :235-248
[3]   THE COMPUTATION OF SCATTER IN MAMMOGRAPHY BY MONTE-CARLO METHODS [J].
DANCE, DR ;
DAY, GJ .
PHYSICS IN MEDICINE AND BIOLOGY, 1984, 29 (03) :237-247
[4]   MONTE-CARLO CALCULATION OF CONVERSION FACTORS FOR THE ESTIMATION OF MEAN GLANDULAR BREAST DOSE [J].
DANCE, DR .
PHYSICS IN MEDICINE AND BIOLOGY, 1990, 35 (09) :1211-1219
[6]  
DANCE DR, 1981, P PATIENT EXPOSURE R, P227
[7]   X-RAY TRANSMISSION FORMULA FOR ANTISCATTER GRIDS [J].
DAY, GJ ;
DANCE, DR .
PHYSICS IN MEDICINE AND BIOLOGY, 1983, 28 (12) :1429-1433
[8]   INFLUENCE OF ANODE AND FILTER MATERIAL ON IMAGE QUALITY AND GLANDULAR DOSE FOR SCREEN FILM MAMMOGRAPHY [J].
DESPONDS, L ;
DEPEURSINGE, C ;
GRECESCU, M ;
HESSLER, C ;
SAMIRI, A ;
VALLEY, JF .
PHYSICS IN MEDICINE AND BIOLOGY, 1991, 36 (09) :1165-1182
[9]   OPTIMIZATION OF SPECTRAL SHAPE IN DIGITAL MAMMOGRAPHY - DEPENDENCE ON ANODE MATERIAL, BREAST THICKNESS, AND LESION TYPE [J].
FAHRIG, R ;
YAFFE, MJ .
MEDICAL PHYSICS, 1994, 21 (09) :1473-1481
[10]   CONTRAST AND DOSE WITH MO-MO, MO-RH, AND RH-RH TARGET-FILTER COMBINATIONS IN MAMMOGRAPHY [J].
GINGOLD, EL ;
WU, XZ ;
BARNES, GT .
RADIOLOGY, 1995, 195 (03) :639-644