Finite element analysis of shear stresses at the implant-bone interface of an acetabular press-fit cup during impingement

被引:15
作者
Voigt, Christian
Kloehn, Carsten
Bader, Rainer
von Salis-Soglio, Georg
Scholz, Roger
机构
[1] Univ Leipzig, Orthopad Klin & Poliklin, D-04103 Leipzig, Germany
[2] HTWK Leipzig FH, Fachbereich Maschinen & Energietech, Markkleeberg, Germany
[3] Univ Rostock, Orthopad Klin & Poliklin, D-2500 Rostock 1, Germany
来源
BIOMEDIZINISCHE TECHNIK | 2007年 / 52卷 / 02期
关键词
finite element analysis; impingement; implant-bone interface; total hip arthroplasty;
D O I
10.1515/BMT.2007.038
中图分类号
R318 [生物医学工程];
学科分类号
0831 [生物医学工程];
摘要
After total hip replacement (THR) impingement of the implant components causes shear stresses at the acetabular implant-bone interface. In the current study the finite element method (FEM) was applied to analyse the shear stresses at a fully bonded implant-bone interface assuming total ingrowth of the cup. The FE model of a press-fit acetabular component and the proximal part of the femoral component incorporates non-linear material and large sliding contact. The model was loaded with a superior-medial joint load of 435 N simulating a two-legged stance. Starting at initial impingement, the femoral component was medially rotated by 20 degrees. The peak tilting shear stress of -2.6 MPa at the impingement site takes effect towards the pole of the cup. The torsional shear stress at the impingement site is zero. On each side of the impingement site, there are extrema of torsional shear stress reaching -1.8 and 1.8 MPa, respectively. The global peak shear stress during impingement may indicate a possible starting point for cup loosening. The pattern of the torsional shear stresses suggests that besides the symmetric lever-out, an additional asymmetrical tilting of the cup occurs that can be explained by the orientation of the applied joint load.
引用
收藏
页码:208 / 215
页数:8
相关论文
共 27 条
[1]
Anderson M J, 1994, J Arthroplasty, V9, P17, DOI 10.1016/0883-5403(94)90133-3
[2]
Experimental analysis of standard, asymmetric and constraint liners for total hip replacement: Investigation of range of motion and stability against dislocation [J].
Bader, R ;
Steinhauser, E ;
Scholz, R ;
Simnacher, M ;
Mittelmeier, W .
ZEITSCHRIFT FUR ORTHOPADIE UND IHRE GRENZGEBIETE, 2004, 142 (05) :577-585
[3]
Method for the evaluation of factors influencing the dislocation stability of total hip endoprotheses [J].
Bader, R ;
Scholz, R ;
Steinhauser, E ;
Busch, R ;
Mittlemeier, W .
BIOMEDIZINISCHE TECHNIK, 2004, 49 (05) :137-144
[4]
Ceramic acetabular cups for total hip replacement - Part 6: Cup design, inclination angle and antetorsion angle have an influence on the range of motion and impingement [J].
Bader, R ;
Willmann, G .
BIOMEDIZINISCHE TECHNIK, 1999, 44 (7-8) :212-219
[5]
Virtual reality computer animation of the effect of component position and design on stability after total hip arthroplasty [J].
Barrack, RL ;
Lavernia, C ;
Ries, M ;
Thornberry, R ;
Tozakoglou, E .
ORTHOPEDIC CLINICS OF NORTH AMERICA, 2001, 32 (04) :569-+
[6]
BEREITER H, 1992, ORTHOPADE, V21, P63
[7]
Hip contact forces and gait patterns from routine activities [J].
Bergmann, G ;
Deuretzbacher, G ;
Heller, M ;
Graichen, F ;
Rohlmann, A ;
Strauss, J ;
Duda, GN .
JOURNAL OF BIOMECHANICS, 2001, 34 (07) :859-871
[8]
Use of constrained acetabular components for hip instability: An average 10-year follow-up study [J].
Bremner, BRB ;
Goetz, DD ;
Callaghan, JJ ;
Capello, WN ;
Johnston, RC .
JOURNAL OF ARTHROPLASTY, 2003, 18 (07) :131-137
[9]
CRIPTON PA, 1993, THESIS KINGSTON ON
[10]
Optimizing acetabular component position to minimize impingement and reduce contact stress [J].
D'Lima, DD ;
Chen, PC ;
Colwell, CW .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 2001, 83A :87-91