Atmospheric pressure photoionisation: An ionization method for liquid chromatography-mass spectrometry

被引:658
作者
Robb, DB
Covey, TR
Bruins, AP
机构
[1] Univ Groningen, Ctr Pharm, NL-9713 AV Groningen, Netherlands
[2] MDS SCIEX, Concord, ON L4K 4V8, Canada
关键词
D O I
10.1021/ac0001636
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
Atmospheric pressure photoionization (APPI) has been successfully demonstrated to provide high sensitivity to LC-MS analysis. A vacuum-ultraviolet lamp designed for photoionization detection in gas chromatography is used as a source of 10-eV photons. The mixture of samples and solvent eluting from an HPLC is fully evaporated prior to introduction into the photoionization region. In the new method, large quantities of an ionizable dopant are added to the vapor generated from the LC eluant, allowing for a great abundance of dopant photoions to be produced. Because the ion source is at atmospheric pressure, and the collision rate is high, the dopant photoions react to completion with solvent and analyte molecules present in the ion source. Using APPI, at an LC flow rate of 200 mu L/min, it is possible to obtain analyte signal intensities 8 times as high as those obtainable with a commercially available corona discharge-atmospheric pressure chemical ionization source.
引用
收藏
页码:3653 / 3659
页数:7
相关论文
共 14 条
[1]   ION MOBILITY DETECTOR FOR GAS-CHROMATOGRAPHY WITH A DIRECT PHOTO-IONIZATION SOURCE [J].
BAIM, MA ;
EATHERTON, RL ;
HILL, HH .
ANALYTICAL CHEMISTRY, 1983, 55 (11) :1761-1766
[2]   ATMOSPHERIC-PRESSURE IONIZATION MASS-SPECTROMETRY [J].
CARROLL, DI ;
DZIDIC, I ;
HORNING, EC ;
STILLWELL, RN .
APPLIED SPECTROSCOPY REVIEWS, 1981, 17 (03) :337-406
[3]   PHOTOIONIZATION DETECTOR FOR OPEN-TUBULAR LIQUID-CHROMATOGRAPHY [J].
DEWIT, JSM ;
JORGENSON, JW .
JOURNAL OF CHROMATOGRAPHY, 1987, 411 :201-212
[4]   TRACE ANALYSIS OF ORGANIC-COMPOUNDS BY HIGH-PERFORMANCE LIQUID-CHROMATOGRAPHY WITH PHOTOIONIZATION DETECTION [J].
DRISCOLL, JN ;
CONRON, DW ;
FERIOLI, P ;
KRULL, IS ;
XIE, KH .
JOURNAL OF CHROMATOGRAPHY, 1984, 302 (OCT) :43-50
[5]  
DRISCOLL JN, 1976, RES DEV, P50
[6]   BENZENE CHARGE-EXCHANGE AT ATMOSPHERIC-PRESSURE FOR LOW-LEVEL DETECTION OF POLLUTANTS IN AMBIENT AIR [J].
KETKAR, SN ;
DULAK, JG ;
DHEANDHANOO, S ;
FITE, WL .
ANALYTICA CHIMICA ACTA, 1991, 245 (02) :267-270
[7]   PHOTO-IONIZATION DETECTOR SENSITIVITY OF ORGANIC-COMPOUNDS [J].
LANGHORST, ML .
JOURNAL OF CHROMATOGRAPHIC SCIENCE, 1981, 19 (02) :98-103
[8]   PHOTOIONIZATION IN AIR WITH ION MOBILITY SPECTROMETRY USING A HYDROGEN DISCHARGE LAMP [J].
LEASURE, CS ;
FLEISCHER, ME ;
ANDERSON, GK ;
EICEMAN, GA .
ANALYTICAL CHEMISTRY, 1986, 58 (11) :2142-2147
[9]   LIQUID-PHASE PHOTO-IONIZATION DETECTOR FOR LIQUID-CHROMATOGRAPHY [J].
LOCKE, DC ;
DHINGRA, BS ;
BAKER, AD .
ANALYTICAL CHEMISTRY, 1982, 54 (03) :447-450
[10]  
Revel'skii I.A., 1985, USSR Patent, Patent No. 1159412