Fatigue fracture of a proximally modular, distally tapered fluted implant with diaphyseal fixation

被引:34
作者
Buttaro, Martin A.
Mayor, Michael B.
Van Citters, Douglas
Piccaluga, Francisco
机构
[1] Italian Hosp Buenos Aires Potosi, Inst Orthopaed Carlos E Ottolenghi, Hip Surg Unit, RA-4215 Buenos Aires, DF, Argentina
[2] Dartmouth Med Sch Senior Orthopaed, Hanover, NH USA
[3] Dartmouth Hitchcock Med Ctr, Hanover, NH USA
[4] Dartmouth Coll, Thayer Sch Engn, Hanover, NH 03755 USA
[5] Dartmouth Coll, Thayer Sch Engn, Ctr Orthopaed, Dartmouth Biomed Engn, Hanover, NH 03755 USA
关键词
modular femoral stem; distally fluted; fracture; complication;
D O I
10.1016/j.arth.2006.07.007
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
We report and analyze the causes of a fracture in a proximally modular, distally tapered fluted MP stem in a 48-year-old woman (168 cm, 67 kg) with severe proximal bone deficiency. Evidence of fatigue failure with striations initiated laterally was observed in the laser etching of the tensile aspect of the prosthesis. However, metallurgical analysis suggested that laser engraving did not alter the microstructure of the stem. Stress due to the absence of proximal femoral bone support may have been sufficiently high to put this particular stem at risk for fatigue fracture. This important complication should be addressed when choosing this therapeutic option in cases with substantial proximal femoral bone loss. Strut allograft support should be recommended in such cases.
引用
收藏
页码:780 / 783
页数:4
相关论文
共 18 条
[1]
Stem modularity is unnecessary in revision total hip arthroplasty - In the affirmative [J].
Barrack, RL .
JOURNAL OF ARTHROPLASTY, 2003, 18 (03) :98-100
[2]
Böhm P, 2001, J BONE JOINT SURG AM, V83A, P1023
[3]
FRACTURE OF FEMORAL PROSTHESIS IN TOTAL HIP-REPLACEMENT ACCORDING TO CHARNLEY [J].
CARLSSON, AS ;
GENTZ, CF ;
STENPORT, J .
ACTA ORTHOPAEDICA SCANDINAVICA, 1977, 48 (06) :650-655
[4]
FRACTURE OF THE FEMORAL COMPONENT AFTER TOTAL HIP-REPLACEMENT - AN ANALYSIS OF 58 CASES [J].
CHAO, EYS ;
COVENTRY, MB .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 1981, 63 (07) :1078-1094
[5]
CHARNLEY J, 1975, CLIN ORTHOP RELAT R, P105
[6]
DUNCAN CP, 1995, INSTR COURSE LECT, V44, P293
[7]
Engelbrecht E, 1987, KLASSIFIKATION BEHAN, P189
[8]
INTERGRANULAR CORROSION-FATIGUE FAILURE OF COBALT-ALLOY FEMORAL STEMS - A FAILURE ANALYSIS OF 2 IMPLANTS [J].
GILBERT, JL ;
BUCKLEY, CA ;
JACOBS, JJ ;
BERTIN, KC ;
ZERNICH, MR .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 1994, 76A (01) :110-115
[9]
RETRIEVAL ANALYSIS OF TOTAL KNEE PROSTHESES - A METHOD AND ITS APPLICATION TO 48 TOTAL CONDYLAR PROSTHESES [J].
HOOD, RW ;
WRIGHT, TM ;
BURSTEIN, AH .
JOURNAL OF BIOMEDICAL MATERIALS RESEARCH, 1983, 17 (05) :829-842
[10]
A modular distal fixation option for proximal bone loss in revision total hip arthroplasty - A 2-to 6-year follow-up study [J].
Kwong, LM ;
Miller, AJ ;
Lubinus, P .
JOURNAL OF ARTHROPLASTY, 2003, 18 (03) :94-97