Strain distribution in plated and unplated sheep tibia - An in vivo experiment

被引:52
作者
Gautier, E [1 ]
Perren, SM [1 ]
Cordey, J [1 ]
机构
[1] Kantonsspital, Dept Orthopaed Surg, CH-1708 Fribourg, Switzerland
来源
INJURY-INTERNATIONAL JOURNAL OF THE CARE OF THE INJURED | 2000年 / 31卷
关键词
plate osteosynthesis; stress shielding; strain distribution; strain gauges; steel plate; titanium plate;
D O I
10.1016/S0020-1383(00)80030-3
中图分类号
R4 [临床医学];
学科分类号
1002 ; 100602 ;
摘要
After plate osteosynthesis changes in bone biology and bone mechanics are observed in the plated bone segment. Compromise of the vascular supply in the plate bed leads to a remodelling process and to a temporary porosity in the bone cortex underneath the plate. In addition, the plate takes over some of the physiological loading of the bone, which in turn alters the normal strain distribution of the cortical bone tissue. The aim of the present study is to determine the tissue deformation of the sheep tibia in vivo and the changes in tissue strain due to plating with plates of different rigidities. Measurements were performed on the intact bone at the mid diaphysis using the strain gauge technique. With different connections on the tension bridge (Wheatstone bridge), the strain was measured separately for pure axial loading, bending, and torsion before and after plating with a 4.5 mm stainless steel or titanium DCP. Under physiological load the sheep tibia is mainly deformed in torsion (62%) and bending (33%), and much less in axial loading (5%). Plating with a steel plate reduces the overall tissue strain by 18%, with a titanium plate by 13%. This reduction is mainly due to a reduction in axial tissue strain due to axial loading and bending and less to the reduction of tissue strain under torsion. In our in vivo model, plating with steel or titanium plates leads to a reduction of the physiological tissue strains. The difference between the different plates is small due to the fact that the high tissue strain under torsion is only slightly affected by plating. Thus, from the purely mechanical paint of view and with regard to preserving normal tissue strains as much as possible, titanium plates offer little advantage compared with stainless steel plates. But, titanium as an implant material may offer advantages with respect to tissue compatibility and infection resistance.
引用
收藏
页码:37 / 44
页数:8
相关论文
共 46 条
[1]   QUANTITATIVE HISTOLOGICAL EVALUATION OF EARLY FRACTURE HEALING OF CORTICAL BONES IMMOBILIZED BY STAINLESS-STEEL AND COMPOSITE PLATES [J].
AKESON, WH ;
WOO, SLY ;
COUTTS, RD ;
MATTHEWS, JV ;
GONSALVES, M ;
AMIEL, D .
CALCIFIED TISSUE RESEARCH, 1975, 19 (01) :27-37
[2]   EFFECTS OF RIGIDITY OF INTERNAL-FIXATION PLATES ON LONG BONE REMODELING - BIOMECHANICAL AND QUANTITATIVE HISTOLOGICAL STUDY [J].
AKESON, WH ;
WOO, SLY ;
RUTHERFORD, L ;
COUTTS, RD ;
GONSALVES, M ;
AMIEL, D .
ACTA ORTHOPAEDICA SCANDINAVICA, 1976, 47 (03) :241-249
[3]  
BRENNWALD J, 1974, HELV CHIR ACTA, V41, P455
[4]  
BRENNWALD J, 1972, LANGENBECKS ARCH C S
[5]   A CUMULATIVE DAMAGE MODEL FOR BONE-FRACTURE [J].
CARTER, DR ;
CALER, WE .
JOURNAL OF ORTHOPAEDIC RESEARCH, 1985, 3 (01) :84-90
[6]   EXPERIMENTAL STUDIES ON THE INFLUENCE OF PLATE MATERIAL ON STRESS-RELIEF AND ATROPHY OF THE BONE BENEATH OSTEOSYNTHESIS PLATES [J].
CLAES, L ;
KINZL, L ;
NEUGEBAUER, R .
BIOMEDIZINISCHE TECHNIK, 1981, 26 (04) :66-71
[7]  
COCHRAN G V B, 1969, Surgical Forum (Chicago), V20, P469
[8]   IMPLANTATION OF STRAIN GAGES ON BONE IN-VIVO [J].
COCHRAN, GVB .
JOURNAL OF BIOMECHANICS, 1972, 5 (01) :119-&
[9]  
CORDEY J, 1987, THESIS U LAUSANNE
[10]  
COUTTS RE, 1976, ORTHOP CLIN N AM, V7, P223