Influence of bubble size distribution on the echogenicity of ultrasound contrast agents -: A study of SonoVue™

被引:340
作者
Gorce, JM [1 ]
Arditi, M [1 ]
Schneider, M [1 ]
机构
[1] Bracco Res SA, CH-1228 Plan Les Ouates, Switzerland
关键词
bubble size distribution; contrast efficacy; SonoVue (TM); ultrasound contrast agents;
D O I
10.1097/00004424-200011000-00003
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
RATIONALE AND OBJECTIVES. TO study the relative contributions of different bubble size classes to SonoVue(TM)'s echogenicity in fundamental acoustic imaging modes. SonoVue(TM) is a contrast agent, previously known as BR1, with a bubble size distribution extending from approximately 0.7 to 10 mum, METHODS. A model for the acoustic response of SonoVue(TM) was determined and validated for a set of experimental data. This model was used to simulate the acoustic response of a standard batch of SonoVue(TM) as the sum of responses of nonoverlapping bubble size classes. RESULTS. The simulation was first validated for a standard SonoVue(TM) bubble size distribution. When this distribution was considered as five size classes with equal numbers of bubbles, it was found that bubbles smaller than 2 mum accounted for 60% of the total number but contained only 5% of the total gas volume. The simulation results indicated marked differences in the acoustic contributions from these classes, with 80% of the acoustic efficacy provided by bubbles 3 to 9 mum in diameter. The study also compared bubble distributions in number, surface, and volume, with the distribution computed in terms of acoustic efficacy, CONCLUSIONS. This study shows why bubble volume is a much better indicator of SonoVue(TM)'s efficacy than is bubble count. A low threshold in diameter was found for SonoVue(TM) microbubbles at approximately 2 mum, under which size bubbles do not contribute appreciably to the echogenicity at medical ultrasound frequencies.
引用
收藏
页码:661 / 671
页数:11
相关论文
共 16 条
[1]   Assessment of myocardial perfusion by intermittent harmonic power Doppler using SonoVue, a new ultrasound contrast agent [J].
Broillet, A ;
Puginier, J ;
Ventrone, R ;
Schneider, M .
INVESTIGATIVE RADIOLOGY, 1998, 33 (04) :209-215
[2]   The measurement of backscatter coefficient from a broadband pulse-echo system: A new formulation [J].
Chen, XC ;
Phillips, D ;
Schwarz, KQ ;
Mottley, JG ;
Parker, KJ .
IEEE TRANSACTIONS ON ULTRASONICS FERROELECTRICS AND FREQUENCY CONTROL, 1997, 44 (02) :515-525
[3]   THE EFFECTS OF AN ELASTIC SOLID-SURFACE LAYER ON THE RADIAL PULSATIONS OF GAS-BUBBLES [J].
CHURCH, CC .
JOURNAL OF THE ACOUSTICAL SOCIETY OF AMERICA, 1995, 97 (03) :1510-1521
[4]   ULTRASOUND SCATTERING PROPERTIES OF ALBUNEX MICROSPHERES [J].
DEJONG, N ;
HOFF, L .
ULTRASONICS, 1993, 31 (03) :175-181
[5]   ABSORPTION AND SCATTER OF ENCAPSULATED GAS FILLED MICROSPHERES - THEORETICAL CONSIDERATIONS AND SOME MEASUREMENTS [J].
DEJONG, N ;
HOFF, L ;
SKOTLAND, T ;
BOM, N .
ULTRASONICS, 1992, 30 (02) :95-103
[6]   HIGHER HARMONICS OF VIBRATING GAS-FILLED MICROSPHERES .1. SIMULATIONS [J].
DEJONG, N ;
CORNET, R ;
LANCEE, CT .
ULTRASONICS, 1994, 32 (06) :447-453
[7]   Acoustic modeling of shell-encapsulated gas bubbles [J].
Frinking, PJA ;
de Jong, N .
ULTRASOUND IN MEDICINE AND BIOLOGY, 1998, 24 (04) :523-533
[8]   Oscillations of polymeric microbubbles: Effect of the encapsulating shell [J].
Hoff, L ;
Sontum, PC ;
Hovem, JM .
JOURNAL OF THE ACOUSTICAL SOCIETY OF AMERICA, 2000, 107 (04) :2272-2280
[9]   Acoustic properties of ultrasonic contrast agents [J].
Hoff, L .
ULTRASONICS, 1996, 34 (2-5) :591-593
[10]  
HOFF L, 1996, IEEE ULTR S SAN ANT, P1441