Mechanical properties and in vitro degradation of bioabsorbable self-expanding braided stents

被引:50
作者
Nuutinen, JP
Clerc, C
Reinikainen, R
Törmälä, P
机构
[1] Tampere Univ Technol, Inst Biomat, FIN-33101 Tampere, Finland
[2] Bionx Implants Inc, St Petersburg 194223, Russia
关键词
bioabsorbable; stent; braiding; PLLA; self-expansion; fibres; in vitro;
D O I
10.1163/156856203763572707
中图分类号
R318 [生物医学工程];
学科分类号
0831 [生物医学工程];
摘要
The aim of this study was to characterize the mechanical and self-expansion properties of braided bioabsorbable stents. In total four different stents were manufactured from PLLA fibres using a braiding technique. The changes in radial pressure stiffness and diameter recovery of the stents were determined initially, and after insertion and release from a delivery device. The braided stents were compared to three commercially available metallic braided stents. The changes in physical and mechanical properties of the PLLA fibres and stents during in vitro degradation were investigated. After release from the delivery device, the PLLA stents did not fully recover to their original diameter. The radial pressure stiffness of the bioabsorbable stents was similar to that of the metallic stents. The in vitro degradation study showed that the stents would keep at least half of their initial radial pressure stiffness for more than 22 weeks.
引用
收藏
页码:255 / 266
页数:12
相关论文
共 27 条
[1]
DEFORMATION CHARACTERISTICS OF A BIOABSORBABLE INTRAVASCULAR STENT [J].
AGRAWAL, CM ;
CLARK, HG .
INVESTIGATIVE RADIOLOGY, 1992, 27 (12) :1020-1024
[2]
Colombo A, 2000, CIRCULATION, V102, P371
[3]
Dunn R.L., 1999, The Encyclopedia of Controlled Drug Delivery, V1, P71
[4]
INVESTIGATION OF STRUCTURE OF SOLUTION GROWN CRYSTALS OF LACTIDE COPOLYMERS BY MEANS OF CHEMICAL-REACTIONS [J].
FISCHER, EW ;
STERZEL, HJ ;
WEGNER, G .
KOLLOID-ZEITSCHRIFT AND ZEITSCHRIFT FUR POLYMERE, 1973, 251 (11) :980-990
[5]
Freeman, 2001, TECH GASTROINTEST EN, V3, P120, DOI DOI 10.1053/TGIE.2001.22150
[6]
GINSBERG G, UNPUB GASTROINTEST E
[7]
Biodegradation of the copolymeric polylactide stent -: Long-term follow-up in a rabbit aorta model [J].
Hietala, EM ;
Salminen, US ;
Ståhls, A ;
Välimaa, T ;
Maasilta, P ;
Törmälä, P ;
Nieminen, MS ;
Harjula, ALJ .
JOURNAL OF VASCULAR RESEARCH, 2001, 38 (04) :361-369
[8]
A STUDY OF THE GEOMETRICAL AND MECHANICAL-PROPERTIES OF A SELF-EXPANDING METALLIC STENT THEORY AND EXPERIMENT [J].
JEDWAB, MR ;
CLERC, CO .
JOURNAL OF APPLIED BIOMATERIALS, 1993, 4 (01) :77-85
[9]
Kangas J, 2001, J BIOMED MATER RES, V58, P121, DOI 10.1002/1097-4636(2001)58:1<121::AID-JBM180>3.0.CO
[10]
2-Z