Decreasing peak flow rate with a new bag-valve-mask device:: effects on respiratory mechanics, and gas distribution in a bench model of an unprotected airway

被引:17
作者
Wagner-Berger, HG
Wenzel, V
Stallinger, A
Voelckel, WG
Rheinberger, K
Stadlbauer, KH
Augenstein, S
Dörges, V
Lindner, KH
Hörmann, C
机构
[1] Univ Innsbruck, Dept Anaesthesiol & Crit Care Med, A-6020 Innsbruck, Austria
[2] Univ Kiel, Dept Anaesthesiol, Kiel, Germany
基金
奥地利科学基金会;
关键词
respiration-artificial; bag-valve ventilation; unprotected-airway; peak inspiratory flow; basic life support; SMART BAG (R);
D O I
10.1016/S0300-9572(03)00032-7
中图分类号
R4 [临床医学];
学科分类号
1002 ; 100602 ;
摘要
Reducing inspiratory flow rate and peak airway pressure may be important in order to minimise the risk of stomach inflation when ventilating an unprotected airway with positive pressure ventilation. The purpose of this study was to assess the effects of a newly developed bag-valve-mask device (SMART BAG(R), O-Two Systems International, Ont., Canada) that limits peak inspiratory flow. A bench model simulating a patient with an unintubated airway was used consisting of a face mask, manikin head, training lung (lung compliance, 100 ml/cm H(2)O, airway resistance 4 cm H(2)O/1/S, lower oesophageal sphincter pressure 20 cm H(2)O and simulated stomach). Twenty nurses were randomised to each ventilate the manikin using a standard single person technique for 1 min (respiratory rate, 12/min) with either a standard adult self-inflating bag, or the SMART BAG(R). The volunteers were blinded to the experimental design of the model until completion of the experimental protocol. The SMART BAG(R) vs. standard self-inflating bag resulted in significantly (P < 0.05) lower mean+/-S.D. peak inspiratory flow rates (32+/-2 vs. 61+/-13 1/min), peak inspiratory pressure (12+/-2 vs. 17+/-2 cm H(2)O), lung tidal volumes (525+/-111 vs. 680+/-154 ml) and stomach tidal volumes (0+/-0 vs. 17+/-36 ml), longer inspiratory times (1.9+/-0.3 vs. 1.5+/-0.3 s), but significantly higher mask leakage (26+/-13 vs. 14+/-8%); mask tidal volumes (700+/-104vs. 785+/-172 ml) were comparable. The mask leakage observed is not an uncommon factor in bag-valve-mask ventilation with leakage fractions of 25-40% having been previously reported. The differences observed between the standard BVM and the SMART BAG(R) are due more to the anatomical design of the mask and the non-anatomical shape of the manikin face than the function of the device. Future studies should remove the mask to manikin interface and should introduce a standardized mask leakage fraction. The use of a two-person technique may have removed the problem of mask leakage. In conclusion, using the SMART BAG(R) during simulated ventilation of an unintubated patient in respiratory arrest significantly decreased inspiratory flow rate, peak inspiratory pressure, stomach tidal volume, and resulted in a significantly longer inspiratory time when compared to a standard self-inflating bag. (C) 2003 Published by Elsevier Science Ireland Ltd.
引用
收藏
页码:193 / 199
页数:7
相关论文
共 18 条
[1]   LOWER ESOPHAGEAL SPHINCTER PRESSURE DURING PROLONGED CARDIAC-ARREST AND RESUSCITATION [J].
BOWMAN, FP ;
MENEGAZZI, JJ ;
CHECK, BD ;
DUCKETT, TM .
ANNALS OF EMERGENCY MEDICINE, 1995, 26 (02) :216-219
[2]   Optimisation of tidal volumes given with self-inflatable bags without additional oxygen [J].
Dörges, V ;
Ocker, H ;
Hagelberg, S ;
Wenzel, V ;
Schmucker, P .
RESUSCITATION, 2000, 43 (03) :195-199
[3]  
FELEGI WB, 1996, ACAD EMERG MED, V3, P441
[4]   TECHNIQUES OF EMERGENCY VENTILATION - A MODEL TO EVALUATE TIDAL VOLUME, AIRWAY PRESSURE, AND GASTRIC INSUFFLATION [J].
JOHANNIGMAN, JA ;
BRANSON, RD ;
DAVIS, K ;
HURST, JM .
JOURNAL OF TRAUMA-INJURY INFECTION AND CRITICAL CARE, 1991, 31 (01) :93-98
[5]   COMPLICATIONS OF CARDIAC RESUSCITATION [J].
KRISCHER, JP ;
FINE, EG ;
DAVIS, JH ;
NAGEL, EL .
CHEST, 1987, 92 (02) :287-291
[6]  
MARTIN PD, 1993, ANAESTHESIA, V48, P133
[7]   CHEST COMPRESSION AND VENTILATION RATES DURING CARDIOPULMONARY-RESUSCITATION - THE EFFECTS OF AUDIBLE TONE GUIDANCE [J].
MILANDER, MM ;
HISCOK, PS ;
SANDERS, AB ;
KERN, KB ;
BERG, RA ;
EWY, GA .
ACADEMIC EMERGENCY MEDICINE, 1995, 2 (08) :708-713
[8]   MEASUREMENT OF VENTILATION DURING CARDIOPULMONARY RESUSCITATION [J].
ORNATO, JP ;
BRYSON, BL ;
DONOVAN, PJ ;
FARQUHARSON, RR ;
JAEGER, C .
CRITICAL CARE MEDICINE, 1983, 11 (02) :79-82
[9]   A NEW NONREBREATHING VALVE [J].
RUBEN, H .
ANESTHESIOLOGY, 1955, 16 (04) :643-645
[10]   COMPLIANCE IN APNEIC ANESTHETIZED ADULTS [J].
SAFAR, P ;
AGUTOESCARRAGA, L .
ANESTHESIOLOGY, 1959, 20 (03) :283-289