Remote electrochemical sensor for monitoring TNT in natural waters

被引:98
作者
Wang, J [1 ]
Bhada, RK
Lu, JM
MacDonald, D
机构
[1] New Mexico State Univ, Dept Chem & Biochem, Las Cruces, NM 88003 USA
[2] New Mexico State Univ, Dept Chem Engn, Las Cruces, NM 88003 USA
关键词
TNT; remote sensor; square-wave voltammetry; environmental monitoring;
D O I
10.1016/S0003-2670(97)00702-2
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
An electrochemical sensor for remote continuous monitoring of the explosive 2,4,6-trinitrotoluene (TNT) is described. The probe relies on rapid and sensitive square-wave voltammetry at a submersible carbon-fiber working-electrode assembly, connected to a 50ft-long shielded cable. Operational conditions have been optimized to meet the specific requirements of remote operations. Tests with untreated river water, groundwater, and drinking water samples, spiked with TNT, displayed results similar to those obtained with synthetic buffer solutions. Due to the low detection potential of TNT, coexisting sample constituents do not contribute to the response. The use of ultrafast square-wave voltammetric scanning, without a regeneration step, leads to a rapid detection of dynamic concentration changes, low ppb detection limits, and high stability. Prospects for using the electrochemical probe for real-time monitoring of TNT are discussed. (C) 1998 Elsevier Science B.V.
引用
收藏
页码:85 / 91
页数:7
相关论文
共 10 条
[1]   DETERMINATION OF NITRO AROMATIC, NITRAMINE, AND NITRATE ESTER EXPLOSIVE COMPOUNDS IN EXPLOSIVE MIXTURES AND GUNSHOT RESIDUE BY LIQUID-CHROMATOGRAPHY AND REDUCTIVE ELECTROCHEMICAL DETECTION [J].
BRATIN, K ;
KISSINGER, PT ;
BRINER, RC ;
BRUNTLETT, CS .
ANALYTICA CHIMICA ACTA, 1981, 130 (02) :295-311
[2]  
CONLEY KA, 1976, N8 US ARM CONSTR ENG
[3]   COMPARISON OF SOLID-PHASE EXTRACTION WITH SALTING-OUT SOLVENT-EXTRACTION FOR PRECONCENTRATION OF NITROAROMATIC AND NITRAMINE EXPLOSIVES FROM WATER [J].
JENKINS, TF ;
MIYARES, PH ;
MYERS, KF ;
MCCORMICK, EF ;
STRONG, AB .
ANALYTICA CHIMICA ACTA, 1994, 289 (01) :69-78
[4]   A displacement flow immunosensor for explosive detection using microcapillaries [J].
Narang, U ;
Gauger, PR ;
Ligler, FS .
ANALYTICAL CHEMISTRY, 1997, 69 (14) :2779-2785
[5]   APPLICATION OF LIQUID-CHROMATOGRAPHY ELECTROCHEMISTRY IN PHARMACEUTICAL AND BIOCHEMICAL-ANALYSIS - A CRITICAL-REVIEW [J].
RADZIK, DM ;
LUNTE, SM .
CRC CRITICAL REVIEWS IN ANALYTICAL CHEMISTRY, 1989, 20 (05) :317-358
[6]   DETECTION OF TNT IN WATER USING AN EVANESCENT-WAVE FIBEROPTIC BIOSENSOR [J].
SHRIVERLAKE, LC ;
BRESLIN, KA ;
CHARLES, PT ;
CONRAD, DW ;
GOLDEN, JP ;
LIGLER, FS .
ANALYTICAL CHEMISTRY, 1995, 67 (14) :2431-2435
[7]   On-site detection of TNT with a portable fiber optic biosensor [J].
ShriverLake, LC ;
Donner, BL ;
Ligler, FS .
ENVIRONMENTAL SCIENCE & TECHNOLOGY, 1997, 31 (03) :837-841
[8]   Renewable-reagent electrochemical sensor for monitoring trace metal contaminants [J].
Wang, J ;
Lu, JM ;
Luo, DB ;
Wang, JY ;
Jiang, M ;
Tian, BM ;
Olsen, K .
ANALYTICAL CHEMISTRY, 1997, 69 (14) :2640-2645
[9]   REMOTE ELECTROCHEMICAL SENSOR FOR TRACE-METAL CONTAMINANTS [J].
WANG, J ;
LARSON, D ;
FOSTER, N ;
ARMALIS, S ;
LU, JM ;
RONGRONG, X ;
OLSEN, K ;
ZIRINO, A .