Validation of partial tissue segmentation of single-channel magnetic resonance images of the brain

被引:58
作者
Grabowski, TJ
Frank, RJ
Szumski, NR
Brown, CK
Damasio, H
机构
[1] Univ Iowa, Coll Med, Dept Neurol, Div Behav Neurol & Cognit Neurosci, Iowa City, IA 52242 USA
[2] Lawrence Livermore Natl Lab, Livermore, CA USA
[3] Univ Iowa, Coll Publ Hlth, Dept Biostat, Iowa City, IA 52242 USA
关键词
D O I
10.1006/nimg.2000.0649
中图分类号
Q189 [神经科学];
学科分类号
071006 ;
摘要
We describe and evaluate a practical, automated algorithm based on local statistical mixture modeling for segmenting single-channel, T1-weighted volumetric magnetic resonance images of the brain into gray matter, white matter, and cerebrospinal fluid. We employed a stereological sampling method to assess, prospectively, the performance of the method with respect to human experts on 10 normal T1-weighted brain scans acquired with a three-dimensional gradient echo pulse sequence. The overall kappa statistic for the concordance of the algorithm with the human experts was 0.806, while that among raters, excluding the algorithm, was 0.802. The algorithm had better agreement with the modal expert decision (kappa = 0.878). The algorithm could not be distinguished from the experts by this measure. We also validated the algorithm oil a simulated MR scan of a digital brain phantom with known tissue composition. Global gray matter and white matter errors were 1% and <1%, respectively, and correlation coefficients with the underlying tissue model were 0.95 for gray matter, 0.98 for white matter, and 0.95 for cerebrospinal fluid. In both approaches to validation, we evaluated both local and global performance of the algorithm. Human experts generated slightly higher global gray matter proportion estimates on the test brain scans relative to the algorithm (3.7%) and on the simulated MR scan relative to the true tissue model (4.4%). The algorithm underestimated gray in some subcortical nuclei which contain admired gray and white matter. We demonstrate the reliability of the method on individual 1 NEX data sets of the test subjects, and its insensitivity to the precise values of initial model parameters. The output of this algorithm is suitable for quantifying cerebral cortical tissue, using a commonly performed commercial pulse sequence. (C) 2000 Academic Press.
引用
收藏
页码:640 / 656
页数:17
相关论文
共 60 条
[1]   Multimodal image coregistration and partitioning - A unified framework [J].
Ashburner, J ;
Friston, K .
NEUROIMAGE, 1997, 6 (03) :209-217
[2]  
Ashburner J., 1998, NEUROIMAGE, V7, DOI [DOI 10.1016/S1053-8119(18)31539-8, 10.1016/S1053-8119(18)31539-8]
[3]   TECHNICAL NOTE - INTENSITY CORRECTION IN SURFACE-COIL MR IMAGING [J].
AXEL, L ;
COSTANTINI, J ;
LISTERUD, J .
AMERICAN JOURNAL OF ROENTGENOLOGY, 1987, 148 (02) :418-420
[4]  
AYLWARD S, 1994, P 3 C VIS BIOM COMP
[5]   INTERACTIVE SEGMENTATION OF CEREBRAL GRAY-MATTER, WHITE-MATTER, AND CSF - PHOTOGRAPHIC AND MR-IMAGES [J].
BARTLETT, TQ ;
VANNIER, MW ;
MCKEEL, DW ;
GADO, M ;
HILDEBOLT, CF ;
WALKUP, R .
COMPUTERIZED MEDICAL IMAGING AND GRAPHICS, 1994, 18 (06) :449-460
[6]   Volumetric analysis of white matter, gray matter, and CSF using fractional volume analysis [J].
Bedell, BJ ;
Narayana, PA .
MAGNETIC RESONANCE IN MEDICINE, 1998, 39 (06) :961-969
[7]   REVIEW OF MR IMAGE SEGMENTATION TECHNIQUES USING PATTERN-RECOGNITION [J].
BEZDEK, JC ;
HALL, LO ;
CLARKE, LP .
MEDICAL PHYSICS, 1993, 20 (04) :1033-1048
[8]   ESTIMATION OF CSF, WHITE AND GRAY-MATTER VOLUMES IN HYDROCEPHALIC CHILDREN USING FUZZY CLUSTERING OF MR-IMAGES [J].
BRANDT, ME ;
BOHAN, TP ;
KRAMER, LA ;
FLETCHER, JM .
COMPUTERIZED MEDICAL IMAGING AND GRAPHICS, 1994, 18 (01) :25-34
[9]   Accuracy and reproducibility of brain and tissue volumes using a magnetic resonance segmentation method [J].
Byrum, CE ;
MacFall, JR ;
Charles, HC ;
Chitilla, VR ;
Boyko, OB ;
Upchurch, L ;
Smith, JS ;
Rajagopalan, P ;
Passe, T ;
Kim, D ;
Xanthakos, S ;
Ranga, K ;
Krishnan, R .
PSYCHIATRY RESEARCH-NEUROIMAGING, 1996, 67 (03) :215-234
[10]  
CHADA RR, 1994, THESIS U IOWA