Pesticide determination by enzymatic inhibition and amperometric detection in a low-temperature cofired ceramics microsystem

被引:22
作者
Llopis, Xavier [1 ]
Ibanez-Garcia, Nuria [1 ]
Alegret, Salvador [1 ]
Alonso, Julian [1 ]
机构
[1] Univ Autonoma Barcelona, Dept Quim Analit, Grp Sensors & Biosensors, Edifici C, E-08193 Barcelona, Spain
关键词
D O I
10.1021/ac062390z
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
Among the several fabrication techniques used to construct microflow systems, the low-temperature cofired ceramics (LTCC) technology, taking advantage of its multilayer approach, is one of the most versatile ones. It permits the integration of several unitary operations of an analytical process in a modular or monolithic way. Moreover, due to its perfect compatibility with screen-printing techniques, it also permits the integration of electronic components used to control the whole system setup. In this work the design, construction, and evaluation of a miniaturized analyzer for pesticide determination that integrates a pretreatment stage, based on two mixers or reactors, and an amperometric detection system to measure the product of an enzymatic inhibition reaction are presented. The detection system was monolithically integrated in the microfluidic platform, and it consisted of a screen-printed reference electrode and two platinum sheets, acting as auxiliary and working electrodes, which were embedded within the ceramic structure. The miniaturized system was characterized and successfully evaluated by determining carbofuran at the nanomolar level.
引用
收藏
页码:3662 / 3666
页数:5
相关论文
共 20 条
[1]  
BARROW GM, 1973, PHYS CHEM, P703
[2]  
Becker H, 2000, ELECTROPHORESIS, V21, P12, DOI 10.1002/(SICI)1522-2683(20000101)21:1<12::AID-ELPS12>3.0.CO
[3]  
2-7
[4]   Enzyme inhibition-based biosensors and biosensing systems: questionable analytical devices [J].
de Castro, MDL ;
Herrera, MC .
BIOSENSORS & BIOELECTRONICS, 2003, 18 (2-3) :279-294
[5]   Microflow injection system based on a multicommutation technique for nitrite determination in wastewaters [J].
del Mar Baeza, Maria ;
Ibanez-Garcia, Nuria ;
Baucells, Joan ;
Bartroli, Jordi ;
Alonso, Julian .
ANALYST, 2006, 131 (10) :1109-1115
[6]   Overview of low temperature co-fired ceramics tape technology for meso-system technology (MsST) [J].
Gongora-Rubio, MR ;
Espinoza-Vallejos, P ;
Sola-Laguna, L ;
Santiago-Avilés, JJ .
SENSORS AND ACTUATORS A-PHYSICAL, 2001, 89 (03) :222-241
[7]   CHARACTERIZATION OF INHIBITORS OF ACETYLCHOLINESTERASE BY AN AUTOMATED AMPEROMETRIC FLOW-INJECTION SYSTEM [J].
GUNTHER, A ;
BILITEWSKI, U .
ANALYTICA CHIMICA ACTA, 1995, 300 (1-3) :117-125
[8]   LTCC meso-analytical system for chloride ion determination in drinking waters [J].
Ibanez-Garcia, N. ;
Goncalves, R. D. Machado ;
da Rocha, Z. Mendes ;
Gongora-Rubio, M. R. ;
Seabra, A. C. ;
Chamarro, J. Alonso .
SENSORS AND ACTUATORS B-CHEMICAL, 2006, 118 (1-2) :67-72
[9]   Continuous flow analytical microsystems based on low-temperature co-fired ceramic technology.: Integrated potentiometric detection based on solvent polymeric ion-selective electrodes [J].
Ibanez-Garcia, Nuria ;
Mercader, Manel Bautista ;
da Rocha, Zaira Mendes ;
Seabra, Carlos Antonio ;
Gongora-Rubio, Mario Ricardo ;
Chamarro, Julian Alonso .
ANALYTICAL CHEMISTRY, 2006, 78 (09) :2985-2992
[10]   EXCHANGEABLE IMMOBILIZED ENZYME REACTOR FOR ENZYME-INHIBITION TESTS IN FLOW-INJECTION ANALYSIS USING A MAGNETIC DEVICE - DETERMINATION OF PESTICIDES IN DRINKING-WATER [J].
KINDERVATER, R ;
KUNNECKE, W ;
SCHMID, RD .
ANALYTICA CHIMICA ACTA, 1990, 234 (01) :113-117