Low-noise two-wired buffer electrodes for bioelectric amplifiers

被引:24
作者
Degen, Thomas
Torrent, Simon
Jaeckel, Heinz
机构
[1] Swiss Fed Inst Technol, Dept Informat Technol & Elect Engn, CH-8001 Zurich, Switzerland
[2] Univ Appl Sci, Haute Ecole Valaisanne, CH-1950 Sion, Switzerland
[3] Swiss Fed Inst Technol, Inst Elect, CH-8001 Zurich, Switzerland
关键词
active electrodes; bioelectric recordings; dc-offset compensation; instrumentation amplifier; two-wired electrodes;
D O I
10.1109/TBME.2006.889781
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Active buffer electrodes are known to improve the immunity of bioelectric recordings against power line interferences. A survey of published work reveals that buffer electrodes are almost exclusively designed using operational amplifiers (opamps). In this paper, we discuss the advantage of utilizing a single transistor instead. This allows for a simple electrode, which is small and requires only two wires. In addition, a single transistor adds considerably less noise when compared to an opamp with the same power consumption. We then discuss output resistance and gain as well as their respective effect on the common mode rejection ratio (CMRR). Finally, we demonstrate the use of two-wired buffer electrodes for a bioelectric amplifier.
引用
收藏
页码:1328 / 1332
页数:5
相关论文
共 8 条
[1]   A pseudodifferential amplifier for bioelectric events with DC-Offset compensation using two-wired amplifying electrodes [J].
Degen, T ;
Jäckel, H .
IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 2006, 53 (02) :300-310
[2]  
Fernandez M, 1997, P IEEE EMBS, V18, P97, DOI 10.1109/IEMBS.1996.656864
[3]  
Ko WH, 1998, P ANN INT IEEE EMBS, V20, P2221, DOI 10.1109/IEMBS.1998.747053
[4]   CLINICAL-APPLICATION OF AN ACTIVE ELECTRODE USING AN OPERATIONAL-AMPLIFIER [J].
NISHIMURA, S ;
TOMITA, Y ;
HORIUCHI, T .
IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 1992, 39 (10) :1096-1099
[5]   MICROELECTRONIC SKIN ELECTRODE [J].
PADMADINATA, FZ ;
VEERHOEK, JJ ;
VANDIJK, GJA ;
HUIJSING, JH .
SENSORS AND ACTUATORS B-CHEMICAL, 1990, 1 (1-6) :491-494
[6]   COMMON-MODE REJECTION RATIO IN DIFFERENTIAL-AMPLIFIERS [J].
PALLASARENY, R ;
WEBSTER, JG .
IEEE TRANSACTIONS ON INSTRUMENTATION AND MEASUREMENT, 1991, 40 (04) :669-676
[7]  
RICHARDSON PC, 1968, AEROSPACE MED, P745
[8]   HIGH-QUALITY RECORDING OF BIOELECTRIC EVENTS .2. LOW-NOISE, LOW-POWER MULTICHANNEL AMPLIFIER DESIGN [J].
VANRIJN, AC ;
PEPER, A ;
GRIMBERGEN, CA .
MEDICAL & BIOLOGICAL ENGINEERING & COMPUTING, 1991, 29 (04) :433-440