The relationship between EAP and EABR thresholds and levels used to program the nucleus 24 speech processor: Data from adults

被引:203
作者
Brown, CJ
Hughes, ML
Luk, B
Abbas, PJ
Wolaver, A
Gervais, J
机构
[1] Univ Iowa, Dept Speech Pathol & Audiol, Iowa City, IA 52242 USA
[2] Univ Iowa, Dept Otolaryngol Head & Neck Surg, Iowa City, IA 52242 USA
关键词
D O I
10.1097/00003446-200004000-00009
中图分类号
R36 [病理学]; R76 [耳鼻咽喉科学];
学科分类号
100104 ; 100213 ;
摘要
Objective: The objective of this study was to determine the relationship between electrically evoked whole nerve action potential (EAP) and electrical auditory brain stem response (EABR) thresholds and MAP threshold (T-level) and maximum comfort level (C-level) for subjects who use the Nucleus 24 cochlear implant system. Design: Forty-four adult Nucleus 24 cochlear implant users participated in this study. EAP thresholds were recorded using the Neural Response Telemetry System developed by Cochlear Corporation. EABR thresholds were measured for a subset of 14 subjects using standard evoked potential techniques. These physiologic thresholds were collected on a set of five electrodes spaced across the cochlea, and were then compared with behavioral measures of T-level and C-level used to program the speech processor. Results: EAP thresholds were correlated with MAP T- and C-levels; however, the correlation was not strong. A technique for improving the correlation by combining measures of T- and C-levels made on one electrode with the EAP thresholds was presented. Correlations between predicted and measured T- and C-levels using this technique were 0.83 and 0.77, respectively. Similar results were obtained using the EABR thresholds for a smaller set of subjects. In general, EABR thresholds were recorded at levels that were approximately 4.7 programming units lower than EAP thresholds. Conclusions: Either EAP or EABR thresholds can be used in combination with a limited amount of behavioral information to predict MAP T- and C-levels with reasonable accuracy.
引用
收藏
页码:151 / 163
页数:13
相关论文
共 21 条
[1]   Summary of results using the nucleus CI24M implant to record the electrically evoked compound action potential [J].
Abbas, PJ ;
Brown, CJ ;
Shallop, JK ;
Firszt, JB ;
Hughes, ML ;
Hong, SH ;
Staller, SJ .
EAR AND HEARING, 1999, 20 (01) :45-59
[2]  
ABBAS PJ, IN PRESS ANN OTOLOGY
[3]  
ABBAS PJ, 1998, 7 S COCHL IMPL CHILD
[4]   ELECTRICALLY EVOKED WHOLE-NERVE ACTION-POTENTIALS - DATA FROM HUMAN COCHLEAR IMPLANT USERS [J].
BROWN, CJ ;
ABBAS, PJ ;
GANTZ, B .
JOURNAL OF THE ACOUSTICAL SOCIETY OF AMERICA, 1990, 88 (03) :1385-1391
[5]   INTRAOPERATIVE AND POSTOPERATIVE ELECTRICALLY-EVOKED AUDITORY BRAIN-STEM RESPONSES IN NUCLEUS COCHLEAR IMPLANT USERS - IMPLICATIONS FOR THE FITTING PROCESS [J].
BROWN, CJ ;
ABBAS, PJ ;
FRYAUFBERTSCHY, H ;
KELSAY, D ;
GANTZ, BJ .
EAR AND HEARING, 1994, 15 (02) :168-176
[6]   Electrically evoked whole nerve action potentials in ineraid cochlear implant users: Responses to different stimulating electrode configurations and comparison to psychophysical responses [J].
Brown, CJ ;
Abbas, PJ ;
Borland, J ;
Bertschy, MR .
JOURNAL OF SPEECH AND HEARING RESEARCH, 1996, 39 (03) :453-467
[7]  
Brown CJ, 1998, AM J OTOL, V19, P320
[8]   Relationship between EABR thresholds and levels used to program the CLARION® speech processor [J].
Brown, CJ ;
Lopez, SM ;
Hughes, ML ;
Abbas, PJ .
ANNALS OF OTOLOGY RHINOLOGY AND LARYNGOLOGY, 1999, 108 (04) :50-57
[9]  
BROWN CJ, 1997, 1997 C IMPL AUD PROS, P96
[10]   HUMAN ACOUSTIC NERVE ACTION POTENTIAL RECORDINGS FROM TYMPANIC MEMBRANE WITHOUT ANESTHESIA [J].
CULLEN, JK ;
ELLIS, MS ;
BERLIN, CI ;
LOUSTEAU, RJ .
ACTA OTO-LARYNGOLOGICA, 1972, 74 (1-2) :15-&