Objective assessment of the anterior tibial translation in Lachman test position: Comparison between three types of measurement

被引:31
作者
Benvenuti J.-F. [1 ,2 ]
Vallotton J.A. [1 ]
Meystre J.-L. [1 ]
Leyvraz P.F. [1 ]
机构
[1] Hop. Orthopedique la Suisse Romande, CH-1005 Lausanne
[2] Swiss Fed. Inst. Technol. Lausanne, PSE
关键词
Anterior cruciate ligament; Arthrometer; Electromagnetic sensing device; Radiological drawer test;
D O I
10.1007/s001670050102
中图分类号
学科分类号
摘要
The purpose of this study was to compare three types of objective measurement in Lachman test position and to validate a new measurement procedure. Twenty-three patients were evaluated after anterior cruciate ligament (ACL) reconstruction at 10 years' mean follow-up. The contralateral knee had a normal ACL. Both knees were compared using Lachman test, arthrometer, radiological drawer, and electromagnetic sensing device (EMS) measurements. The values were obtained by two trained orthopaedic surgeons independently. Arthrometer assessment was done with the Kneelax 3 (Biodex) at 88 and 132 N. Radiological drawer was measured with a load of 10 kg. EMS is composed of an electromagnetic transmitter and two sensors. Displacement of the sensors was measured along the three perpendicular axes of the transmitter. Data after processing give the displacement of the sensors perpendicular to the tibial axis. The load was applied on the calf perpendicular to the tibial axis with a 10-kg counterweight. Metallic objects must be avoided around the set-up. No significant difference between examiners was observed in clinical (P = 0.45) or objective (P > 0.3) measurements. The best correlation was obtained between arthrometer and EMS measurements (R = 0.94). Arthrometric and radiological drawer were poorly correlated (R = 0.65). The values of the medial radiological drawer were quantitatively similar to the 132 N arthrometric values. The sensitivity of EMS to skin movements was tested. The correlation between the EMS drawer and the other movements was not significant (R ≤ 0.37). Quantitative correspondence with Lachman grades was only 36% of the arthrometer and EMS values and 37% of the radiological measurements. The effectiveness of EMS to evaluate anterior tibial displacement in Lachman test position after ACL reconstruction is demonstrated in this study. This method is non-invasive and comfortable for the patient, but needs a careful set-up. This method can potentially be used to assess rotational laxities.
引用
收藏
页码:215 / 219
页数:4
相关论文
共 15 条
[1]
Anderson C., Gillquist J., Instrumented testing for evaluation of sagittal knee laxity, Clin Orthop, 256, pp. 178-184, (1990)
[2]
Benvenuti J.F., Holm M., Rakotomanana L., Pioletti D.P., Leyvraz P.F., Invariant description of rotations: Applications to the in vitro kinematics of the knee, Arch Physiol Biochem, 104, (1996)
[3]
Butler D.L., Noyes F.R., Grood E.S., Ligamentous restraints to anterior-posterior drawer in the human knee. A biomechanical study, J Bone Joint Surg, 62, pp. 259-270, (1980)
[4]
Daniel D.M., Malcom L.L., Losse G., Instrumented measurement of anterior laxity of the knee, J Bone Joint Surg, 67, pp. 720-725, (1985)
[5]
Edixhofen P., Huiskes R., De Graaf R., Van Rens J.G., Sloof T.J., Accuracy and reproducibility of instrumented knee-drawer tests, J Orthop Res, 5, pp. 378-387, (1987)
[6]
Karrholm J., Selvic G., Elmqvist L.G., Three-dimensional instability of the anterior cruciate deficient knee, J Bone Joint Surg, 70, pp. 777-783, (1988)
[7]
Kowalk D.L., Wojtys E.M., Disher J., Quantitative analysis of the measuring capabilities of the KT-1000 knee ligament arthrometer, Am J Sports Med, 21, pp. 744-747, (1993)
[8]
Lerat J.L., Moyen B., Jenny J.Y., Perrier J.P., A comparison of pre-operative evaluation of anterior knee laxity by dynamic X-rays and by the arthrometer KT 1000, Knee Surg Sports Traumatol Arthrose, 1, pp. 54-59, (1993)
[9]
Noyes F.R., Cummings J.F., Grood E.S., The diagnosis of knee motion limits, subluxations, and ligament injury, Am J Sports Med, 19, pp. 163-171, (1991)
[10]
Pearcy M.J., Hindle R.J., New method for the non-invasive three-dimensional measurement of human back movement, Clin Biomech, 4, pp. 73-79, (1989)