A fiber-optic twin sensor for dose measurements in radiation therapy

被引:4
作者
Harald Bueker
Friedrich W. Haesing
Frank Pfeiffer
Hermann-Joseph Schmitz
机构
[1] ZEL-EGS,Forschungszentrum Jülich GmbH
[2] ProCom GmbH,undefined
关键词
fiber-optic sensors; dose measurement; ionizing radiation; in; dosimetry; in; sensors; radiation therapy;
D O I
10.1007/BF02936010
中图分类号
学科分类号
摘要
The developed fiber-optic twin sensor allowsin- vivo dose measurement in the tissue of the patient’s body. It consists of two radiation-sensitive fibers. The radiation-induced attenuations of these fibers are used to determine the dose absorbed in tissue. The two signals have a linear dose response, but they depend differently on the energy of the ionizing radiation. Evaluation of the two sensor signals permits a nearly tissue-equivalent dose indication.