The mechanical role of laminar hook protection of pedicle screws at the caudal end vertebra

被引:10
作者
Margulies J.Y. [1 ]
Casar R.S. [2 ]
Caruso S.A. [3 ]
Neuwirth M.G. [2 ]
Haher T.R. [3 ]
机构
[1] Spine Service, Department of Orthopedic Surgery, Montefiore Medical Center, New York, NY 10467
[2] Department of Orthopedic Surgery, Hospital for Joint Diseases, New York, NY
[3] Department of Orthopedic Surgery, St. Vincent's Medical Center, New York, NY
关键词
Laminar hooks; Pedicle screws; Spinal instrumentation; Torsional instability;
D O I
10.1007/BF01322446
中图分类号
学科分类号
摘要
Biomechanical studies have shown hooks to be superior to pedicle screws in pull-out, especially in osteoporosis. This study evaluates the possible increase in stiffness of pedicle screws provided by laminar hooks while applying non-destructive forces to a vertebrectomy model assembled with Compact Cotrel Dubousset (CCD) instrumentation. Synthetic vertebrae were employed in a three-level vertebrectomy model. CCD screw-based three-level constructs with and without sublaminar hooks in the caudal element were tested in flexion, extension, compression, lateral bending, and torsion. There was no statistically significant advantage in adding inferior laminar hooks to a caudal end vertebra that had bilateral pedicle screws in any of the testing modes. Torsional stability, however, was augmented, but not significantly. Torsional instability and osteoporotic bone may be the clinical justifications for adding laminar hooks below screws in the caudal end vertebra.
引用
收藏
页码:245 / 248
页数:3
相关论文
共 18 条
[1]  
Provisional standard test method for static and dynamic spinal implants assembly in a corpectomy model, Annual Book of ASTM Standards, (1994)
[2]  
Ashman R.B., Galpin R.D., Corin J.D., Johnston C.E., Biomechanical analysis of pedicle screw instrumentation systems in a corpectomy model, Spine, 14, pp. 1398-1405, (1989)
[3]  
Carl A.L., Tromanhauser S.G., Roger D.J., Pedicle instrumentation for thoracolumbar burst fractures and fracture dislocations, Spine, 17, SUPPL., pp. 317-324, (1992)
[4]  
Chopin D., Morin C., CotrelDubousset instrumentation for adolescent idiopathic scoliosis, Textbook of Spinal Surgery, (1991)
[5]  
Coe J.D., Warden K.E., Herzig M.A., McAfee P.C., Influence of bonemineral density on the fixation of thoracolumbar implants. A comparative study of transpedicular screws, laminar hooks and spinous process wires, Spine, 15, pp. 902-907, (1990)
[6]  
Cunnigham B.W., Softer J.C., Shono Y.S., McAfee M.D., Static and cyclical biomechanical analysis of pedicle screw spinal constructs, Spine, 18, pp. 1677-1688, (1993)
[7]  
Ebelke D.K., Asher M.A., Neff J.R., Kraker D.P., Survivorship analysis of VSP spine instrumentation in the treatment of thoraeolumbar and lumbar burst fractures, Spine, SUPPL., pp. 428-432, (1991)
[8]  
Engler G.L., Cotrel-Duboussel in strumentation for reduction of fracture dislocations of the spine, J Spinal Disord, 3, pp. 62-66, (1990)
[9]  
Esses S.I., Sachs B.L., Dreyzin V., Complications associated with the technique of pedicle screw fixation. A selected survey of ABS members, Spine, 18, pp. 2238-2239, (1993)
[10]  
Farcy J.P., Weidenbaum M., Michelsen C.B., Hoeltzel D.A., Athanasiou K., Comparative biomechanical study of spinal fixation using CotrelDuboussel instrumentation, Spine, 12, pp. 877-881, (1987)