The accuracy of portable abdominal ultrasound equipment in measuring postvoid residual volume

被引:41
作者
Alnaif B. [1 ]
Drutz H.P. [1 ,2 ]
机构
[1] Mount Sinai Hospital, Toronto, Ont.
[2] Section of Urogynecology, Mount Sinai Hospital, 600 University Avenue, Toronto
关键词
Bladder scan; Bladder volume; Catheterization; Portable abdominal ultrasound; Postvoid residual; Prospective study;
D O I
10.1007/s001920050048
中图分类号
学科分类号
摘要
This study examines the accuracy of a new portable abdominal ultrasound machine (Bladder Scan, BVI 2500) used to measure postvoid urine residual (PVR). Using this machine, we started prospectively measuring residual urine in the first 80 women undergoing uroflowmetry in our urodynamic unit. Ultrasound PVR measurements were done immediately prior to catheterization while the patient was in the supine position. The catheterized postvoid residual was used as the gold standard. Eighty paired measurements were done on 78 women. Compared to catheterized PVR, ultrasound PVR measurements tend to underestimate and correlated poorly with the actual residual volume. The reading was considered accurate if it was within 25% of the catheterized PVR. The ultrasound PVR measurements were most accurate (60.6%) when the readings were below 50 ml and least accurate (10%) when readings were higher than 150 ml. Readings between 50 and 150 ml were 27% accurate. A measurement of zero '000' was common (44%), usually reflecting volumes of less than 50 ml (84% of cases). However, it could indicate that the bladder had been missed altogether. Partial measurement of the bladder volume, where the lateral bladder borders are missing, produces readings much below the actual volume: a 'tip of the iceberg' phenomenon. We therefore advocate caution when interpreting PVR measurements from portable abdominal ultrasound machines, and if an accurate measurement of PVR is necessary, catheterization remains a more reliable method.
引用
收藏
页码:215 / 218
页数:3
相关论文
共 10 条
[1]  
Meares, E.M., Current patterns on nosocomial urinary tract infection (1991) Urology, 36 (SUPPL.), pp. 9-12
[2]  
Cardenas, D.D., Kelly, E., Krieger, J.N., Chapman, W.H., Residual urine volume in patients with spinal cord injury: Measurement with a portable ultrasound instrument (1988) Arch Phys Med Rehab, 69, pp. 514-516
[3]  
Massagli, T.L., Cardenas, D.D., Kelly, E.W., Experience with portable ultrasound equipment and measurement of urine volumes: Inter-user reliability and factors of patient position (1989) J Urol, 142, pp. 969-971
[4]  
Mainprize, T.C., Drutz, H.P., Accuracy of total bladder volume and residual urine measurements: Comparison between real-time ultrasonography and catheterization (1989) Am J Obstet Gynecol, 160, pp. 1013-1016
[5]  
Griffiths, C.J., Murray, A., Ramsden, P.D., Accuracy and repeatability of bladder volume measurement using ultrasonic imaging (1986) J Urol, 136, pp. 808-812
[6]  
Alfthan, O., Mattsson, T., Ultrasound method of measuring residual urine (1969) Ann Chir Gyn Fenniae, 58, pp. 300-303
[7]  
Regeth, J.C., Langer, K., Ultrasonic assessment of residual urine volume (1982) Urol Tes, 10, pp. 57-60
[8]  
Ireton, R.C., Krieger, J.N., Cardenas, D.D., Bladder volume determination using a dedicated, portable ultrasound scanner (1990) J Urol, 143, pp. 909-911
[9]  
Ouslander, J.G., Simmons, S., Tuico, E., Use of a portable ultrasound device to measure post-void residual volume among incontinent nursing home residents (1994) J Am Geriatr Soc, 42, pp. 1189-1192
[10]  
Bent, A.E., Nahhas, D.E., McLennan, M.T., Portable ultrasound determination of urinary residual volume (1997) Int Urogynecol J, 8, pp. 200-202