Optimization of Fizeau-based optical coherence tomography with a reference Michelson interferometer

被引:19
作者
V. M. Gelikonov
G. V. Gelikonov
P. A. Shilyagin
机构
[1] Russian Academy of Sciences,Institute of Applied Physics
关键词
Optical Coherence Tomography; Spectral Component; Shot Noise; Optical Coherence Tomography Image; Excess Noise;
D O I
10.3103/S106287380801022X
中图分类号
学科分类号
摘要
The method of Fizeau-based optical coherence tomography with a reference Michelson interferometer has been optimized. The real-system noise and dynamic range have been estimated. The time parameters of the system are optimized according to the proposed noise model. A good correspondence between the numerical estimates and experimental data is shown.
引用
收藏
页码:93 / 97
页数:4