Reliability of pedicle screw assessment utilizing plain radiographs versus CT reconstruction

被引:64
作者
Berlemann U. [1 ]
Heini P. [1 ]
Müller U. [1 ]
Stoupis C. [2 ]
Schwarzenbach O. [1 ]
机构
[1] Department of Orthopaedic Surgery, Inselspital, Univsersity of Bern, CH-3010 Bern, Freiburgstrasse
[2] Department of Diagnostic Radiolgoy, Inselspital, University of Bern, Bern
关键词
Accuracy; CT reconstruction; Pedicle screw; Placement; Plain radiographs;
D O I
10.1007/BF01834069
中图分类号
学科分类号
摘要
The position of 119 pedicle screws was assessed on plain antero- posterior and lateral radiographs taken immediately post-operatively and at 3 months' follow-up. The readings of five independent observers were compared with the 'gold standard' of CT reconstructions. The position of only 41% of implants (range 14%-56%) was assessed correctly on the plain radiographs (47% on follow-up films). Two-thirds of CT-detectable perforations were missed. As shown with perforations of the anterior cortex, detectability increased significantly with magnitude of perforation. No specifically difficult anatomic level or direction of malplacement could be identified. Interobserver variation was considerable. Plain radiographs were shown to be of limited use in assessing the position of pedicle screws.
引用
收藏
页码:406 / 410
页数:4
相关论文
共 19 条
[1]  
Berlemann U., Monin D., Arm E., Nolte L.P., Ozdoha C., Planning and insertion of pedicle screws with computer assistance, J Spinal Disord, 10, pp. 117-124, (1997)
[2]  
Bernard T.N., Seibert C.E., Pedicle diameter determined by computed tomouraphy, Spine, 17, (1992)
[3]  
Castro W.H.M., Halm H., Jerosch J., Malms J., Steinbeck J., Blasius S., Accuracy of pedicle screw placement in lumbar vertebrae, Spine, 21, pp. 1320-1324, (1996)
[4]  
Esses S.I., Sachs B.L., Dreyzin V., Complications associated with the technique of pedicle screw fixation, Spine, 18, pp. 2231-2239, (1993)
[5]  
Farber G.L., Place H.M., Mazur R.A., Casey Jones D.E., Damiano T.R., Accuracy of pedicle screw placement in lumbar fusions by plain radiographs and computed tomography, Spine, 20, pp. 1494-1499, (1995)
[6]  
Gertzbein S., Robbins S., Accuracy of transpedicular screw placement in vivo, Spine, 15, pp. 11-14, (1990)
[7]  
Haaker R.G., Eickhoff U., Schopphoff E., Steffen R., Jergas M., Kramer J., Verification of the position of pedicle screws in lumbar spinal fusion, Eur Spine J, 6, pp. 125-128, (1997)
[8]  
Krag M.H., Weaver D.L., Beynnon B.D., Huugh L.D., Morphometry of the thoracic and lumbar spine related to transpedicular screw placement for surgical spine fixation, Spine, 13, pp. 27-32, (1988)
[9]  
Learch T., Massie J.B., Puthria M., Ahlgren B., Garfin S.R., Evaluation of pedicle screw placement utilizing plain x-rays and computed tomography, Proceedings of the International Society for the Study of the Lumbar Spine, 23rd Meeting, (1996)
[10]  
Misenheimer G.R., Peek R.D., Wiltse L.L., Rothman S.L.G., Widell E.H., Anatomic analysis of pedicle cortical and cancellous diameter as related to screw size, Spine, 14, pp. 367-372, (1989)