Closed-loop control of the inspired fraction of oxygen in mechanical ventilation

被引:15
作者
Tehrani F. [1 ]
Rogers M. [2 ]
Lo T. [3 ]
Malinowsksi T. [2 ]
Afuwape S. [2 ]
Lum M. [2 ]
Grundl B. [2 ]
Terry M. [2 ]
机构
[1] Coll. of Eng. and Computer Science, California State Univ. Fullerton, Dept. of Electrical Engineering, Fullerton, CA 92834-6870
[2] Department of Respiratory Care, Loma Linda University, Medical Center, Loma Linda, CA
[3] Medical Intensive Care Unit, Loma Linda University, Medical Center, Loma Linda, CA
关键词
Closed-loop control; Mechanical ventilation; Oxygen;
D O I
10.1023/A:1024261021473
中图分类号
学科分类号
摘要
Objective. Supplemental oxygen treatment of patients on mechanical ventilation is crucial in maintaining the patients' oxygen levels in the normal range. The purpose of this study was to evaluate the effectiveness of a closed-loop controller for automatic adjustment of the fraction of inspired oxygen, FIO2. More specifically, the aim of the study was to assess the robustness of the controller in correcting hypoxemia as well as its effectiveness in prevention of hyperoxemia and oxygen toxicity. Methods. The microprocessor-based feedback control system combines a rapid control algorithm with a proportional-integral-derivative (PID) control procedure to automatically adjust FIO2. The system is designed to prevent hypoxemia by applying a stepwise control procedure in response to rapid declines in arterial oxygen saturation while fine-tuning FIO2 and avoiding hyperoxemia by resuming to the PID control procedure when appropriate. The system includes a sophisticated safeguard unit which is designed to communicate any oxygenation problems or measurement artifacts to the medical personnel while keeping FIO2 at a safe and sufficiently high level. The control system has been tested by using computer simulations as well as animal studies. Results. In response to different disturbances, the arterial oxygen saturation returned to the normal safe range within less than 20 seconds, thereby avoiding any untoward effects of hypoxemia. Under steady state conditions, the variations in arterial oxygen saturation were kept within ±3% of the mean value. The controller corrected hypoxemia within seconds while preventing hyperoxemia, rejecting artifacts, and minimizing exposure to high concentrations of oxygen. Conclusion. The results of the study attest to the reliability of the proposed closed-loop control scheme for automatic adjustment of FIO2. Further evaluation of the controller will require testing the effectiveness of the system on different patient groups.
引用
收藏
页码:367 / 376
页数:9
相关论文
共 17 条
[1]  
Mitamura Y., Mikami T., Yamamoto K., Mimura K., A dual control system for assisting respiration, Med Biol Eng, 13, 6, pp. 846-854, (1975)
[2]  
Beddis I.R., Collins P., Levy N.M., Godfrey S., Silverman M., A new technique for servo-control of arterial oxygen tension in preterm infants, Arch Dis Child, 54, pp. 278-280, (1979)
[3]  
Sano A., Kikuchi M., Adaptive control of arterial oxygen pressure of newborn infants under incubator oxygen treatments, Proc IEE, 132, 5 PART D, pp. 205-211, (1985)
[4]  
Yu C., He W.G., So J.M., Roy R., Kaufman H., Newell J.C., Improvement in arterial oxygen control using multiple model adaptive control procedures, IEEE Trans Biomed Eng, BME-34, 8, pp. 567-574, (1987)
[5]  
Morozoff P.E., Evans R.W., Closed-loop control of SaO<sub>2</sub> in the neonate, Biomed Instr Tech, 26, pp. 117-123, (1992)
[6]  
Tehrani F.T., A microcomputer oxygen control system for ventilatory therapy, Ann Biomed Eng, 20, 5, pp. 547-558, (1992)
[7]  
Tehrani F.T., Bazar A.R., A feedback controller for supplemental oxygen treatment of newborn infants: A simulation study, Med Eng Phys, 16, pp. 329-333, (1994)
[8]  
Raemer D.B., Ji X., Topulos G.P., FI<sub>X</sub> controller: An instrument to automatically adjust inspired oxygen fraction using feedback control from a pulse oximeter, J Clin Monit, 13, pp. 91-101, (1997)
[9]  
Tehrani F.T., A control system for oxygen therapy of premature infants, Proc IEEE Eng Med Biol Conf, 23, 2, pp. 2059-2062, (2001)
[10]  
Solimano A.J., Smyth J.A., Mann T.K., Albersheim S.G., Lockitch G., Pulse oximetry advantages in infants with bronchopulmonary dysplasia, Pediatrics, 78, 5, pp. 844-849, (1986)