Low aseptic loosening and revision rate in Zweymüller-Plus total hip arthroplasty with ceramic-on-ceramic bearings

被引:7
作者
Bouras T. [1 ]
Repantis T. [1 ]
Fennema P. [2 ]
Korovessis P. [1 ]
机构
[1] Orthopaedic Department, General Hospital “Agios Andreas”, Tsertidou 1, Patras
[2] Institute for Health Sciences, Rotterdam
关键词
Aseptic loosening; Ceramic-on-ceramic; Metal-on-metal; Zweymüller-Plus THA;
D O I
10.1007/s00590-013-1314-y
中图分类号
学科分类号
摘要
Abstract: A recent comparative study reported lower revision rate and higher survival in ceramic-on-ceramic (CoC) than in metal-on-metal (MoM) coupling in Zweymüller-Plus THA. We retrospectively studied 87 consecutive patients (94 hips), who primarily received Zweymüller-Plus THA (SL-Plus stem, ceramic 28-mm ball head, Bicon cup) with third-generation CoC bearing surfaces for osteoarthritis. Survival, revision and osteolysis rate were studied and compared to those observed in historical series from this institution with the same THA and MoM articulation. The mean ± SD follow-up was 7.4 ± 1.7 years (range 6–10 years). Four hips (4 %) in four patients were revised 1–6 years following primary implantation: one (1 %) for aseptic loosening of SL-Plus stem and Bicon 5 years after index surgery; one (1 %) for malpositioning of the Bicon and recurrent dislocations 1 year following implantation; and two (2 %) for septic loosening 1 and 6 years postoperatively. Intraoperatively on inspection, there were neither findings for impingement in the revised hips, nor fracture of the Bicon ceramic inlay and ball head. Expansile osteolysis was shown around SL-Plus stem and Bicon in the revised for aseptic loosening hip. With revision of any component for aseptic loosening as the endpoint, the survival was 99 % (95 % CI 91–100 %), while for revision for any reason, it was 95 % (95 % CI 87–98 %) at 10 years. Zweymüller-Plus THA with CoC bearings showed, at mid-term follow-up, high survival and low revision rate for aseptic loosening. These rates were better than the historical series of the same THA with MoM articulation performed by the same surgeon.; Level of evidence: Level IV. © 2013, Springer-Verlag France.
引用
收藏
页码:1439 / 1445
页数:6
相关论文
共 30 条
[1]
Petsatodis G.E., Et al., Primary cementless total hip arthroplasty with an alumina ceramic-on-ceramic bearing: results after a minimum of twenty years of follow-up, J Bone Joint Surg Am, 92, pp. 639-644, (2010)
[2]
Korovessis P., Repantis T., Zafiropoulos A., High medium-term survivorship and durability of Zweymüller-Plus total hip arthroplasty, Arch Orthop Trauma Surg, 131, 5, pp. 603-611, (2011)
[3]
Korovessis P., Petsinis G., Repanti M., Repantis T., Metallosis after contemporary metal-on-metal total hip arthroplasty. Five to nine-year follow-up, J Bone Joint Surg Am, 88, pp. 1183-1191, (2006)
[4]
Milosev I., Trebse R., Kovac S., Cor A., Pisot V., Survivorship and retrieval analysis of Sikomet metal-on-metal total hip replacements at a mean of seven years, J Bone Joint Surg Am, 88, pp. 1173-1182, (2006)
[5]
Repantis T., Vitsas V., Korovessis P., Poor mid-term survival of the low-carbide metal-on-metal Zweymüller-plus total hip arthroplasty system: a concise follow-up, at a minimum of ten years, of a previous report*, J Bone Joint Surg Am, 95, 6, pp. e331-e334, (2013)
[6]
Milosev I., Kovac S., Trebse R., Levasic V., Pisot V., Comparison of ten-year survivorship of hip prostheses with use of conventional polyethylene, metal-on-metal, or ceramic-on-ceramic bearings, J Bone Joint Surg Am, 94, pp. 1756-1763, (2012)
[7]
Capello W.N., D'Antonio J.A., Feinberg J.R., Manley M.T., Naughton M., Ceramic-on-ceramic total hip arthroplasty: update, J Arthroplasty, 23, pp. 39-43, (2008)
[8]
Yeung E., Et al., Mid-term results of third-generation alumina-on-alumina ceramic bearings in cementless total hip arthroplasty: a ten-year minimum follow-up, J Bone Joint Surg Am, 94, pp. 138-144, (2012)
[9]
Lee Y.-K., Et al., Alumina-on-alumina total hip arthroplasty: a concise follow-up, at a minimum of ten years, of a previous report, J Bone Joint Surg Am, 92, pp. 1715-1719, (2010)
[10]
Lusty P.J., Et al., Third-generation alumina-on-alumina ceramic bearings in cementless total hip arthroplasty, J Bone Joint Surg Am, 89, pp. 2676-2683, (2007)