Correlation of Narcotrend Index, entropy measures, and spectral parameters with calculated propofol effect-site concentrations during induction of propofol-remifentanil anaesthesia

被引:12
作者
Grouven U. [1 ,2 ]
Beger F.A.
Schultz B. [1 ]
Schultz A. [1 ]
机构
[1] Department of Anaesthesionology, Klinikum Hannover Oststadt, Hannover Medical School, D-30659 Hannover
[2] Medizinische Informatik/Biometrie, Klinikum Hannover Oststadt, Medizinische Hochschule Hannover, D-30659 Hannover
关键词
EEG monitoring; Entropy; Narcotrend; Propofol; Spectral parameters;
D O I
10.1007/s10877-005-2917-6
中图分类号
学科分类号
摘要
Objective: The aim of this study was to compare the EEG parameter Narcotrend Index with the spectral and entropy-based EEG parameters median frequency, 95% spectral edge frequency, burst-compensated 95% spectral edge frequency, spectral entropy, amplitude entropy, and approximate entropy with regard to their ability to describe cerebral anaesthetic drug effects during induction of propofol-remifentanil anaesthesia. Methods. Three induction schemes were studied with 10 patients each receiving 2 mg propofol/kg/60s (group 1), 4 mg/kg/120s (group 2), and 4 mg/kg/240s (group 3). The EEG was recorded with the EEG monitor Narcotrend®. To analyse the relation between drug effect and EEG parameters, Spearman rank correlation of the different EEG parameters with the calculated propofol effect-site concentration was computed. Results. In all groups Narcotrend Index showed the highest correlation with the propofol effect-site concentration and the lowest variability of individual correlation values. Furthermore, only the Narcotrend Index showed a monophasic behaviour over the entire time period analysed. In the group of entropy parameters approximate entropy yielded the best results. Among the spectral parameters the burst-compensated 95% spectral edge frequency had the highest correlation with the propofol effect-site concentration. It was markedly higher than for the standard spectral edge frequency. The correlations of median frequency and amplitude entropy with propofol effect-site concentration were the lowest. Conclusions. Changes in the propofol effect-site concentration during induction of anaesthesia were best described by the multivariate Narcotrend Index compared to conventional spectral EEG parameters and different entropy measures. © Springer Science + Business Media, Inc. 2005.
引用
收藏
页码:231 / 240
页数:9
相关论文
共 42 条
[1]  
Heier T., Steen P.A., Assessment of anaesthesia depth, Acta Anaesthesiol. Scand., 40, pp. 1087-1100, (1996)
[2]  
Kreuer S., Molter G., Biedler A., Larsen R., Schoth S., Wilhelm W., Narcotrend stages and end-tidal desflurane concentrations - An investigation during recovery from desflurane/remifentanil anaesthesia, Anaesthesist, 51, pp. 800-804, (2002)
[3]  
Kreuer S., Biedler A., Larsen R., Altmann S., Wilhelm W., Narcotrend™ monitoring allows faster emergence and a reduction of drug consumption in propofol-remifentanil anesthesia, Anesthesiology, 99, pp. 34-41, (2003)
[4]  
Munte S., Munte T.F., Grotkamp J., Et al., Implicit memory varies as a function of hypnotic electroencephalogram stage in surgical patients, Anesth. Analg., 97, pp. 132-138, (2003)
[5]  
Schultz B., Grouven U., Schultz A., Automatic classification algorithms of the EEG monitor Narcotrend for routinely recorded EEG data from general anaesthesia: A validation study, Biomed. Tech., 47, pp. 9-13, (2002)
[6]  
Wilhelm W., Kreuer S., Larsen R., Narcotrend EEG monitoring during total intravenous anaesthesia in 4.630 patients, Anaesthesist, 51, pp. 980-988, (2002)
[7]  
Schwilden H., Stoeckel H., Quantitative EEG analysis during anaesthesia with isoflurane in nitrous oxide at 1.3 and 1.5 MAC, Br. J. Anaesth., 59, pp. 738-745, (1987)
[8]  
Schwender D., Daunderer M., Mulzer S., Klasing S., Finsterer U., Peter K., Spectral edge frequency of the electroencephalogram to monitor depth of anaesthesia with isoflurane or propofol, Br. J. Anaesth., 77, pp. 179-184, (1996)
[9]  
Jansen B.H., Nonlinear dynamics and quantitative EEG analysis, Electroenceph. Clin. Neurophysiol., 45, SUPPL., pp. 39-56, (1996)
[10]  
Pijn J.P., Neerven J.V., Noest A., Lopes da Silva F.H., Chaos or noise in EEG signals