A novel phantom technique for evaluating the performance of PET auto-segmentation methods in delineating heterogeneous and irregular lesions

被引:13
作者
Berthon B. [1 ]
Marshall C. [1 ]
Holmes R. [2 ]
Spezi E. [3 ]
机构
[1] Wales Research and Diagnostic Positron Emission Tomography Imaging Centre, Cardiff University - PETIC, room GF705 Ground floor ‘C’ Block, Heath Park, Cardiff
[2] Department of Medical Physics and Bioengineering, University Hospitals Bristol, Bristol
[3] School of Engineering, Cardiff University, Cardiff, Wales
关键词
!sup]18[!/sup]F-fluorodeoxyglucose; Image segmentation; Imaging phantoms; Inkjet printing; Positron emission tomography; Radiotherapy;
D O I
10.1186/s40658-015-0116-1
中图分类号
学科分类号
摘要
Background: Positron Emission Tomography (PET)-based automatic segmentation (PET-AS) methods can improve tumour delineation for radiotherapy treatment planning, particularly for Head and Neck (H&N) cancer. Thorough validation of PET-AS on relevant data is currently needed. Printed subresolution sandwich (SS) phantoms allow modelling heterogeneous and irregular tracer uptake, while providing reference uptake data. This work aimed to demonstrate the usefulness of the printed SS phantom technique in recreating complex realistic H&N radiotracer uptake for evaluating several PET-AS methods. Methods: Ten SS phantoms were built from printouts representing 2mm-spaced slices of modelled H&N uptake, printed using black ink mixed with 18F-fluorodeoxyglucose, and stacked between 2mm thick plastic sheets. Spherical lesions were modelled for two contrasted uptake levels, and irregular and spheroidal tumours were modelled for homogeneous, and heterogeneous uptake including necrotic patterns. The PET scans acquired were segmented with ten custom PET-AS methods: adaptive iterative thresholding (AT), region growing, clustering applied to 2 to 8 clusters, and watershed transform-based segmentation. The difference between the resulting contours and the ground truth from the image template was evaluated using the Dice Similarity Coefficient (DSC), Sensitivity and Positive Predictive value. Results: Realistic H&N images were obtained within 90 min of preparation. The sensitivity of binary PET-AS and clustering using small numbers of clusters dropped for highly heterogeneous spheres. The accuracy of PET-AS methods dropped between 4% and 68% for irregular lesions compared to spheres of the same volume. For each geometry and uptake modelled with the SS phantoms, we report the number of clusters resulting in optimal segmentation. Radioisotope distributions representing necrotic uptakes proved most challenging for most methods. Two PET-AS methods did not include the necrotic region in the segmented volume. Conclusions: Printed SS phantoms allowed identifying advantages and drawbacks of the different methods, determining the most robust PET-AS for the segmentation of heterogeneities and complex geometries, and quantifying differences across methods in the delineation of necrotic lesions. The printed SS phantom technique provides key advantages in the development and evaluation of PET segmentation methods and has a future in the field of radioisotope imaging. © 2015, Berthon et al.
引用
收藏
页码:1 / 17
页数:16
相关论文
共 32 条
[1]  
Nestle U., Kremp S., Schaefer-Schuler A., Sebastian-Welsch C., Hellwig D., Ru C., Comparison of different methods for delineation of 18F-FDG PET-positive tissue for target volume definition in radiotherapy of patients with non-small cell lung cancer, J. Nucl. Med., 46, 8, pp. 1342-1348, (2005)
[2]  
Hatt M., Cheze le Rest C., Descourt P., Dekker A., De Ruysscher D., Oellers M., Lambin P., Pradier O., Visvikis D., Accurate automatic delineation of heterogeneous functional volumes in positron emission tomography for oncology applications, Int. J. Radiat. Oncol. Biol. Phys., 77, 1, pp. 301-308, (2010)
[3]  
Cheebsumon P., Yaqub M., van Velden F.H.P., Hoekstra O.S., Lammertsma A.A., Boellaard R., Impact of [18F]FDG PET imaging parameters on automatic tumour delineation: need for improved tumour delineation methodology, Eur. J. Nucl. Med. Mol. Imaging, 38, pp. 2136-2144, (2011)
[4]  
Le Maitre A., Hatt M., Pradier O., Cheze-le Rest C., Visvikis D., Impact of the accuracy of automatic tumour functional volume delineation on radiotherapy treatment planning, Phys. Med. Biol., 57, 17, pp. 5381-5397, (2012)
[5]  
Performance measurements of scintillation cameras, Washington DC. Standards Publication No NU1, (2001)
[6]  
Siegel J.A., Benedetto A.R., Jaszczak R.J., Lancaster J.L., Madsen M.T., Woodsen W.W., Zimmerman R.E., Rotating scintillation camera SPECT acceptance testing and quality control, AAPM report, 22, (1987)
[7]  
Berthon B., Marshall C., Evans M., Spezi E., Evaluation of advanced automatic PET segmentation methods using non-spherical thin-wall inserts, Med. Phys, 41, 2, (2014)
[8]  
Turkington T.G., Degrado T.R., Sampson W.H., Small spheres for lesion detection phantoms, in, IEEE Nucl. Sci. Symp. Conf. Rec, 4, pp. 2234-2237, (2001)
[9]  
Bazanez-Borgert M., Bundschuh R., Herz M., Martinez M.-J., Schwaiger M., Ziegler S.I., Radioactive spheres without inactive wall for lesion simulation in PET, Z. Med. Phys, 18, 1, pp. 37-42, (2008)
[10]  
Drever L., Roa W., McEwan A., Robinson D., Comparison of three image segmentation techniques for target volume delineation in positron emission tomography, J. Appl. Clin. Med. Phys, 8, 2, pp. 93-109, (2007)