Temperature monitoring in newborns using thermospot

被引:13
作者
Pejaver R.K. [1 ,2 ]
Nisarga R. [1 ]
Gowda B. [1 ]
机构
[1] Division of Neonatology, Department of Pediatrics, Kempegowda Inst. of Medical Sciences, Bangalore
[2] Vijayanagar, Bangalore 560040
关键词
Hypothermia; Newborns; Temperature monitoring;
D O I
10.1007/BF02730715
中图分类号
学科分类号
摘要
Objective: Hypothermia is a common problem in newborns. This is more so in community environment especially in lowbirth weight and preterm babies. Methods: A simple liquid crystal temperature monitoring device was used in 20 infants on a continuous basis to monitor for hypothermia. This was compared with the measurement of rectal temperature using a rectal thermometer. The colour changes and measures to be taken when hypothermia occurs was explained to the mothers and was well understood by all. Results: The Thermo Spot was in agreement with the rectal temperature in 310 out of 313 measurements. This includes 299 times when temperature was normal; and 11 times when the temperature was hypothermic. It underestimated the temperature of three infants, falsely indicating hypothermia in a normothermic child with normal body temperature. It correctly identified hypothermia in every case. Conclusion: ThermoSpot device is a simple accurate device allowing continuous thermal monitoring of low birth weight infants, especially in resource poor setting.
引用
收藏
页码:795 / 796
页数:1
相关论文
共 11 条
[1]  
Vhatson K., Fant M.E., Cloherty J.P., Edn.
[2]  
Mann T.P., Elliot R.I.K., Neonatal cold injury due to accidental exposure to cold, Lancet, 1, pp. 229-234, (1957)
[3]  
Devi P.S., Et al., Neonatal hypothermia in the community, Proc Annual Conv NNF India, (2002)
[4]  
Morley D., Blumenthal I., A new hypothermia indicator, Lancet, 355, 9204, pp. 659-660, (2000)
[5]  
Rutter N., Temperature control and its disorders, Edn.
[6]  
Cutter J., Recording patient temperature - Are we getting it right?, Professional Nurse, 9, pp. 608-616, (1994)
[7]  
Kennedy N., Gondwe L., Morley D.C., Temperature monitoring with ThermoSpot in Malawi, Lancet, 355, 9212, (2000)
[8]  
Kambarami R., Chidede O., Pereira N., Thermospot in the detection of neonatal hypothermia, Ann Trop Pediatr, 22, pp. 219-223, (2002)
[9]  
Blumenthal I., Should we ban the mercury thermometer?, J Roy Soc Med, 85, pp. 553-555, (1992)
[10]  
Glass D.C., Hall A.J., Harrington J.M., Control of Substances Hazardous to Health. Guidance to Initial Assessment in Hospitals, (1989)