RADIOSURGERY TARGET POINT ALIGNMENT ERRORS DETECTED WITH PORTAL FILM VERIFICATION

被引:21
作者
SERAGO, CF
LEWIN, AA
HOUDEK, PV
GONZALEZARIAS, S
SCHWADE, JG
ABITOL, A
MARCIALVEGA, V
机构
[1] UNIV MIAMI,DEPT RADIAT ONCOL,MIAMI,FL 33152
[2] BAPTIST HOSP MIAMI,MIAMI STEREOTACT RADIOSURG CTR,MIAMI,FL
[3] BAPTIST HOSP MIAMI,CTR CANC TREATMENT,MIAMI,FL
来源
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS | 1992年 / 24卷 / 04期
关键词
RADIOSURGERY; RADIOTHERAPY; QUALITY ASSURANCE;
D O I
10.1016/0360-3016(92)90728-Z
中图分类号
R73 [肿瘤学];
学科分类号
100214 ;
摘要
Stereotactic radiosurgery with a linear accelerator requires an accurate match of the therapeutic radiation distribution to the localized target volume. Techniques for localization of the target volume using CT scans and / or angiograms have been described. Alignment of the therapeutic radiation distribution to the intended point in stereotactic space is usually accomplished using precision mechanical scales which attach to the head ring. The present work describes a technique used to verify that the stereotactic coordinates of the center of the intended radiation distribution are in agreement with the localized target point coordinates. This technique uses anterior / posterior and lateral accelerator portal verification films to localize the stereotactic coordinates of the center of the radiation distribution with the patient in the treatment position. The results of 26 cases have been analyzed. Alignment errors of the therapeutic radiation distribution in excess of 1 mm have been found using the portal film verification procedure.
引用
收藏
页码:777 / 780
页数:4
相关论文
共 15 条
[1]   STEREOTACTIC RADIOSURGERY WITH THE LINEAR-ACCELERATOR - TREATMENT OF ARTERIOVENOUS-MALFORMATIONS [J].
BETTI, OO ;
MUNARI, C ;
ROSLER, R .
NEUROSURGERY, 1989, 24 (03) :311-321
[2]   EXTERNAL STEREOTACTIC IRRADIATION BY LINEAR-ACCELERATOR [J].
COLOMBO, F ;
BENEDETTI, A ;
POZZA, F ;
AVANZO, RC ;
MARCHETTI, C ;
CHIEREGO, G ;
ZANARDO, A .
NEUROSURGERY, 1985, 16 (02) :154-160
[3]   STEREOTAXIC SINGLE HIGH-DOSE RADIATION-THERAPY OF BENIGN INTRACRANIAL MENINGIOMAS [J].
ENGENHART, R ;
KIMMIG, BN ;
HOVER, KH ;
WOWRA, B ;
STURM, V ;
VANKAICK, G ;
WANNENMACHER, M .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1990, 19 (04) :1021-1026
[4]   CEREBRAL RADIATION SURGERY USING MOVING FIELD IRRADIATION AT A LINEAR-ACCELERATOR FACILITY [J].
HARTMANN, GH ;
SCHLEGEL, W ;
STURM, V ;
KOBER, B ;
PASTYR, O ;
LORENZ, WJ .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1985, 11 (06) :1185-1192
[5]  
HEIBURN MP, 1983, INSTRUCTION MANUAL B
[6]   STEREOTAXIC RADIOTHERAPY TECHNIQUE FOR SMALL INTRACRANIAL LESIONS [J].
HOUDEK, PV ;
FAYOS, JV ;
VANBUREN, JM ;
GINSBERG, MS .
MEDICAL PHYSICS, 1985, 12 (04) :469-472
[7]   BRAGG-PEAK PROTON-BEAM THERAPY FOR ARTERIOVENOUS-MALFORMATIONS OF THE BRAIN [J].
KJELLBERG, RN ;
HANAMURA, T ;
DAVIS, KR ;
LYONS, SL ;
ADAMS, RD .
NEW ENGLAND JOURNAL OF MEDICINE, 1983, 309 (05) :269-274
[8]  
LEKSELL L, 1971, STEREOTAXIS RADIOSUR
[9]   A SYSTEM FOR STEREOTACTIC RADIOSURGERY WITH A LINEAR-ACCELERATOR [J].
LUTZ, W ;
WINSTON, KR ;
MALEKI, N .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1988, 14 (02) :373-381
[10]   HEAVY CHARGED-PARTICLE STEREOTACTIC RADIOSURGERY - CEREBRAL-ANGIOGRAPHY AND CT IN THE TREATMENT OF INTRACRANIAL VASCULAR MALFORMATIONS [J].
PHILLIPS, MH ;
FRANKEL, KA ;
LYMAN, JT ;
FABRIKANT, JI ;
LEVY, RP .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1989, 17 (02) :419-426