FACTORS THAT INFLUENCE THE PREVALENCE OF POSITIVE CATCH TRIALS IN GLAUCOMA PATIENTS

被引:12
作者
REYNOLDS, M
STEWART, WC
SUTHERLAND, S
机构
[1] MED UNIV S CAROLINA,STORM EYE INST,171 ASHLEY AVE,CHARLESTON,SC 29425
[2] MED UNIV S CAROLINA,DEPT BIOMETRY,CHARLESTON,SC 29425
[3] MED UNIV S CAROLINA,DEPT OPHTHALMOL,CHARLESTON,SC 29425
关键词
D O I
10.1007/BF00920059
中图分类号
R77 [眼科学];
学科分类号
100212 ;
摘要
We evaluated factors that influence the catchtrial response in automated perimetry by consecutively examining 408 glaucoma patients (703 eyes) with either the program 30-2 or a 76-point screen with a quantified defect on the Humphrey Field Analyzer. The prevalence of positive catch trials using both programs was: 15.9% fixation losses, 5.5% false-positives, and 17.0% false-negatives. The only significant difference between the two programs was a lower prevalence of false-negatives on the program 30-2 (P < 0.01). False-positive responses showed a significant positive correlation to fixation losses, and the catch trial response in general was correlated significantly to defect depth, test time, age, and visual acuity (P < 0.05). This study suggests that the prevalence of positive catch trials is not an independent variable when testing glaucoma patients with automated perimetry and is not reduced by using a screening strategy over a full-threshold exam. © 1990 Springer-Verlag.
引用
收藏
页码:338 / 341
页数:4
相关论文
共 9 条
[1]  
BICKLERBLUTH M, 1989, OPHTHALMOLOGY, V96, P616
[2]  
ENGER C, 1987, ARCH OPHTHALMOL-CHIC, V105, P1355
[3]   SOME ASPECTS OF AUTOMATION OF PERIMETRY [J].
FANKHAUSER, F ;
SPAHR, J ;
BEBIE, H .
SURVEY OF OPHTHALMOLOGY, 1977, 22 (02) :131-141
[4]  
FANKHAUSER F, 1983, PERIMETER DIGEST, P31
[5]   CHANGES IN DIFFERENTIAL THRESHOLD IN PATIENTS WITH GLAUCOMA DURING PROLONGED PERIMETRY [J].
HEIJL, A ;
DRANCE, SM .
BRITISH JOURNAL OF OPHTHALMOLOGY, 1983, 67 (08) :512-516
[6]  
KATZ J, 1988, ARCH OPHTHALMOL-CHIC, V106, P1252
[7]   FULL THRESHOLD VERSUS QUANTIFICATION OF DEFECTS FOR VISUAL-FIELD TESTING IN GLAUCOMA [J].
STEWART, WC ;
SHIELDS, MB ;
OLLIE, AR .
GRAEFES ARCHIVE FOR CLINICAL AND EXPERIMENTAL OPHTHALMOLOGY, 1989, 227 (01) :51-54
[8]  
WHALEN WR, 1985, COMPUTERIZED VISUAL, P85
[9]  
1986, OWNERS MANUAL HUMPHR, P6