IMPACT OF THE NEW CUFF DESIGN ON RELIABILITY OF THE AS800 ARTIFICIAL URINARY SPHINCTER

被引:79
作者
LIGHT, JK
REYNOLDS, JC
机构
关键词
URINARY INCONTINENCE; ARTIFICIAL ORGANS;
D O I
10.1016/S0022-5347(17)37319-6
中图分类号
R5 [内科学]; R69 [泌尿科学(泌尿生殖系疾病)];
学科分类号
1002 ; 100201 ;
摘要
The effect of 2 cuff design changes on the mechanical reliability of the current AS800 artificial urinary sphincter was assessed in 126 patients. The surface-treated cuff was introduced in 1983 and the narrow-back design was introduced in 1987. Mean followup for the surface-treated cuff was 40 months, while that for the narrow back was 27.2 months. The incidence of cuff leaks was 1.3%, while the revision rate for clinically significant pressure atrophy, in the absence of a bladder flap urethroplasty, was 2.3%. The overall revision rate for clinically significant pressure atrophy was 9%. No leaks involving the balloon or tubing occurred. The mechanical reliability of the current AS800 artificial urinary sphincter has improved significantly.