FINITE-ELEMENT MODELING OF FEMORAL-SHAFT FRACTURE FIXATION TECHNIQUES POST TOTAL HIP-ARTHROPLASTY

被引:49
作者
MIHALKO, WM [1 ]
BEAUDOIN, AJ [1 ]
CARDEA, JA [1 ]
KRAUSE, WR [1 ]
机构
[1] VIRGINIA COMMONWEALTH UNIV,MED COLL VIRGINIA,DIV ORTHOPED SURG,RICHMOND,VA 23298
关键词
D O I
10.1016/0021-9290(92)90087-H
中图分类号
Q6 [生物物理学];
学科分类号
071011 [生物物理学];
摘要
The presence of a femoral prosthesis superior to a shaft fracture severely complicates fixation and treatment. This study uses two-dimensional, multithickness, plane stress finite-element models of a femur with prosthesis to investigate the stresses developed with the application of three popular fixation techniques: revision to a long stem prosthesis, lateral plating with a cortical bone allograft strut and cerclage wires, and custom plate application with proximal Parham band fixation with distal cortical screws (Ogden plate). The plate and bone contact as well as the fracture site contact were modelled by using orthotropic elements with custom-fit moduli so that only the normal stress to the interface was significant. A thermal analogy was used to model the cerclage and Parham band preloads so that representative preloads in the proximal fixation of the two types of plate treatments could be modelled. A parametric study was performed with the long-prosthesis model to show variations in stem lengths of one, two and three femoral diameters distal to the fracture site. The Ogden plate model showed a transfer of tensile stress near the proximalmost band, with the highest tensile stress being at the fracture site with evidence of stress shielding of the proximal lateral cortex. The cortical bone strut model showed a transfer of tensile stress to the bone strut but showed less shielding of the proximal cortex. The cerclage wires at the base of the bone strut showed the highest changes in load with the distalmost wire increasing to almost four times its original preload. The increase in stem length for the long-prosthesis model showed stress shielding around the fracture site with a transfer of stress occurring more distally with each increment in stem length.
引用
收藏
页码:469 / 476
页数:8
相关论文
共 17 条
[1]
A QUANTITATIVE ASSESSMENT OF COMPRESSION PLATE FIXATION INVIVO - AN EXPERIMENTAL-STUDY USING THE SHEEP RADIUS [J].
BAGGOTT, DG ;
GOODSHIP, AE ;
LANYON, LE .
JOURNAL OF BIOMECHANICS, 1981, 14 (10) :701-711
[2]
BARRINGTON TW, 1984, COMPLICATIONS TOTAL, P30
[3]
BEAUDOIN AJ, 1989, T ORTHOP RES SOC, V14, P406
[4]
FRACTURES OF THE FEMUR IN RELATION TO CEMENTED HIP PROSTHESES [J].
COOKE, PH ;
NEWMAN, JH .
JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 1988, 70 (03) :386-389
[5]
TELEMETRIC FORCE MEASUREMENTS ACROSS THE HIP AFTER TOTAL ARTHROPLASTY [J].
DAVY, DT ;
KOTZAR, GM ;
BROWN, RH ;
HEIPLE, KG ;
GOLDBERG, VM ;
HEIPLE, KG ;
BERILLA, J ;
BURSTEIN, AH .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 1988, 70A (01) :45-50
[6]
HUISKES R, 1986, T ORTHOP RES SOC, V11, P466
[7]
HUISKES R, 1980, ACTA ORTHOP SCAND S, V189, P1
[8]
FRACTURE OF THE IPSILATERAL FEMUR IN PATIENTS WITH TOTAL HIP-REPLACEMENT [J].
JOHANSSON, JE ;
MCBROOM, R ;
BARRINGTON, TW ;
HUNTER, GA .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 1981, 63 (09) :1435-1442
[9]
MIHALKO WM, 1990, UNPUB EVALUATION OGD
[10]
Ogden WS, 1978, ORTHOP T, V2, P70