IMPLEMENTATION AND TEST OF A SYNCHROTRON RADIATION POSITION MONITOR AT A USER BEAM LINE

被引:3
作者
BRODSKY, EL
KLEMAN, KJ
ROGERS, G
RIOUX, D
PATEL, R
HOCHST, H
机构
[1] Synchrotron Radiation Center, University of Wisconsin-Madison, Stoughton
关键词
D O I
10.1063/1.1142694
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
A slotted plate photodiode beam position monitor was built and installed on a user beam line. The detector is positioned in front of the first optical component of the beam line intercepting approximately 3 mrad of synchrotron radiation. A 25-mu-m-thick Be foil strongly attenuates photons with h-nu < 800 eV and improves the position sensitivity by reducing the vertical angular divergence of the effective photon beam by more than a factor of 10. With a 125-mu-m gap at 45-degrees angle between the diode plates, the detector has a linear working range of approximately +/- 800-mu-m. The vertical position sensitivity of the beam position monitor is approximately 5-mu-m.
引用
收藏
页码:519 / 522
页数:4
相关论文
共 7 条
[2]   UHV PHOTOELECTRON X-RAY-BEAM POSITION MONITOR [J].
JOHNSON, ED ;
OVERSLUIZEN, T .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 1990, 291 (1-2) :427-430
[4]  
KRINSKY S, 1983, HDB SYNCHROTRON RA A, V1, P65
[5]  
LAMBERTSON G, 1987, AIP C P, V153, P1413
[6]  
MITSUHASHI T, 1987, US PARTICLE ACCEL C, P576
[7]   A PHOTON-BEAM POSITION AND ANGLE SHIFT MEASUREMENT SYSTEM FOR ALADDIN [J].
YIN, Y ;
BARANIAK, T .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 1986, 246 (1-3) :413-416