INTRAOPERATIVE RASTER PHOTOGRAMMETRY THE PAR CORNEAL TOPOGRAPHY SYSTEM

被引:20
作者
BELIN, MW
机构
[1] Division of Corneal and External Disease and Refractive Surgery, Albany Medical College, Department of Ophthalmology, Albany, New York
关键词
ERBIUM YAG LASER; PHOTOABLATION; PHOTOGRAMMETRY; PHOTOREFRACTION; RASTERSTEREOGRAPHY; TOPOGRAPHY; VIDEOKERATOSCOPE;
D O I
10.1016/S0886-3350(13)80405-1
中图分类号
R77 [眼科学];
学科分类号
100212 [眼科学];
摘要
The PAR Corneal Topography System (CTS) is a computer-driven corneal imaging system that uses close-range raster photogrammetry to measure and produce a topographic map of the corneal surface. The CTS determines distortion in a projected two-dimensional grid. Unlike Placido-disc-based videokeratoscopes, the PAR CTS produces a true topographic map (elevation map) and requires neither a smooth reflective surface nor precise spatial alignment for accurate imaging. Because the system uses two noncoaxial optical paths, it can be integrated into other optical devices. A modified CTS was integrated into an experimental erbium: YAG photoablative laser. The CTS successfully imaged corneas before, after, and during laser photoablation. Its ability to image nonreflective surfaces and to be integrated into other optical systems may make it suitable for intraoperative refractive monitoring.
引用
收藏
页码:188 / 192
页数:5
相关论文
共 12 条
[1]
Warnicki JW, Rehkopf PG, Curtin SA, Et al., Corneal topography using computer analyzed rasterstereographic images, Appl Opt, 27, pp. 1135-1140, (1988)
[2]
Belin MW, Litoff D, Strods SJ, Et al., The PAR Technology Corneal Topography System, Refract Corneal Surg, 8, pp. 88-96, (1992)
[3]
Litoff D, Belin MW, Winn SS, Smith RS, PAR Technology Corneal Topography System. ARVO Abstracts, Invest Ophthalmol Vis Sci, 32, 4, (1991)
[4]
Hannush SB, Crawford SL, Waring GO, Et al., Accuracy and precision of keratometry, photokeratoscopy, and corneal modeling on calibrated steel balls, Arch Ophthalmol, 107, pp. 1235-1239, (1989)
[5]
Hannush SB, Crawford SL, Waring GO, Et al., Reproducibility of normal corneal power measurements with a keratometer, photokeratoscope, and video imaging system, Arch Ophthalmol, 108, pp. 539-544, (1990)
[6]
Belin MW, Zloty P, Accuracy of the Par Corneal Topography System with spatial misalignment, CLAD J, 19, pp. 64-68, (1993)
[7]
Koch DD, Wakil JS, Samuelson SW, Haft EA, Comparison of the accuracy and reproducibility of the keratometer and the EyeSys Corneal Analysis System Model 1, J Cataract Refract Surg, 18, pp. 342-347, (1992)
[8]
Tsilimbaris MK, Vlachonikolis IG, Siganos D, Et al., Comparison of keratometric readings as obtained by javal ophthalmometer and corneal analysis system (EyeSys), Refract Corneal Surg, 7, pp. 368-373, (1991)
[9]
Arffa RC, Warnicki JW, Rehkopf PG, Corneal topography using rasterstereography, Refract Corneal Surg, 5, pp. 414-417, (1989)
[10]
El Hage SG, A computerized corneal topographer for use in refractive surgery, Refract Corneal Surg, 5, pp. 418-424, (1989)