ISOTHERM MAPPING IN HYPERTHERMIA USING SUBTRACTION X-RAY COMPUTED-TOMOGRAPHY

被引:42
作者
BENTZEN, SM
OVERGAARD, J
JORGENSEN, J
机构
关键词
D O I
10.1016/S0167-8140(84)80066-3
中图分类号
R73 [肿瘤学];
学科分类号
100214 ;
摘要
The prospects for the use of X-ray computed tomography (CT) for isotherm mapping were investigated. The spatial distribution of CT changes was obtained by digital subtraction of images taken before and during heating of a 200 mm diameter cylindrical 4% agar phantom. Due to a rather unfavorable signal-to-noise ratio, the noise level in the resulting image had to be reduced by low-pass filtering using the 2-dimensional Fast Fourier Transform followed by spatial averaging. The resulting isotherms showed a resolution of 0.5.degree. C at a spatial resolution of 4.2 mm. Isotherms so obtained are presented and the possibility of using the method in clinical hyperthermia [for cancer therapy in human] is discussed.