IMPROVING THE APPLICABILITY OF MYOCARDIAL DENSITOMETRY AND PARAMETRIC IMAGING BY EXTENDED AUTOMATED DENSOGRAM ANALYSIS

被引:7
作者
LANG, M [1 ]
BRENNECKE, R [1 ]
HAUDE, M [1 ]
RENNEISEN, U [1 ]
ERBEL, R [1 ]
MEYER, J [1 ]
机构
[1] UNIV MAINZ,MED CLIN 2,D-55101 MAINZ,GERMANY
来源
INTERNATIONAL JOURNAL OF CARDIAC IMAGING | 1995年 / 11卷 / 02期
关键词
MYOCARDIAL PERFUSION; QUALITY CONTROL; PARAMETRIC IMAGING; X-RAY DENSITOMETRY;
D O I
10.1007/BF01844708
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
In clinical applications the analysis of X-ray contrast densograms acquired in regions of interest (ROI's) over the myocardium is disturbed by many complex factors. For this reason we acquire redundant densogram information for quality control before extracting densitometric parameters. In our approach, initially some stable measures of quality for densograms are used to lower the influence of poor quality densograms by a quality weighted averaging. For example a shape quality measure, Q(1), is calculated using regions of optimal and minimal acceptable quality defined with respect to a prototype densogram. Not a few myocardial ROI's yield densograms that differ from single-source densograms (SSD's) due to e.g. superposition of different perfusion beds or the position of the ROI relative to the coronary sinus or stenoses. This might result in a densogram shape with oscillating or plateau behavior. For densograms of a such general shape many parameters defined in the usual way do not depend smoothly on the densogram values. The conventional definitions of some parameters (appearance time, rise time) are therefore extended for application to multi-maxima densograms as well as to SSD's. These new methods are evaluated using digitized clinical angiocardiograms and are applied to parametric imaging (pixeldensograms) in a slightly modified way. Taking into account the densogram quality, its shape and its origin results in a considerable improvement both for densitometry and parametric imaging of myocardial perfusion.
引用
收藏
页码:105 / 115
页数:11
相关论文
共 16 条
[1]  
BHARGAVA V, 1991, COMPUT CARDIOL, P129
[2]  
BRENNECKE R, 1982, COMPUT CARDIOL, P143
[3]  
CUSMA JT, 1991, COMPUT CARDIOL, P125
[4]   DETERMINATION OF CORONARY FLOW RESERVE BY DIGITAL ANGIOGRAPHY - VALIDATION OF A PRACTICAL METHOD NOT REQUIRING POWER INJECTION OR ELECTROCARDIOGRAPHIC GATING [J].
GURLEY, JC ;
NISSEN, SE ;
ELION, JL ;
BOOTH, DC ;
DEMARIA, AN .
JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 1990, 16 (01) :190-197
[5]  
HAHNE HJ, 1985, HERZ, V10, P220
[6]  
HAHNE HJ, 1979, BIOMED TECHNIK, V24, P191
[7]  
Haude M, 1990, Int J Card Imaging, V5, P183, DOI 10.1007/BF01833987
[8]  
HAUDE M, 1991, COMPUT CARDIOL, P141
[9]  
Pijls N H, 1990, Int J Card Imaging, V5, P191, DOI 10.1007/BF01833988
[10]  
Simon R, 1990, Int J Card Imaging, V5, P203, DOI 10.1007/BF01833989