NONINVASIVE BLOOD-PRESSURE MONITORING AT THE FINGER FOR STUDYING SHORT LASTING PRESSOR-RESPONSES IN MAN

被引:23
作者
CHRISTEN, Y [1 ]
WAEBER, B [1 ]
NUSSBERGER, J [1 ]
BRUNNER, HR [1 ]
机构
[1] CHU VAUDOIS,DIV HYPERTENS,CH-1011 LAUSANNE,SWITZERLAND
关键词
D O I
10.1002/j.1552-4604.1990.tb03631.x
中图分类号
R9 [药学];
学科分类号
1007 ;
摘要
The study of vasoactive agents in man often requires accurate measurement of short-lasting changes in blood pressure. Using a noninvasive photoplethysmographic device (Finapres), the authors investigated in normotensive subjects whether rapid increases in blood pressure can be assessed precisely by monitoring finger blood pressure continuously. Six volunteers were studied on two consecutive days. On the first day, increasing doses of angiotensin I were injected intravenously with the aim to find a test dose which raised systolic blood pressure by 25 to 40 mm Hg. After oral administration of a placebo, the same test dose was injected repeatedly over the next 24 hours. On the second day, the subjects took either 6.25 (n=3) or 25 mg (n=3) captopril PO and the serial administration of the test dose of angiotensin I was continued for the next 4 hours. After placebo intake there was a good reproducibility of the blood pressure response to angiotensin I with a coefficient of variation of 15 ± 4.5% (Mean ± SD, n=6). Captopril caused a dose-dependent inhibition of the pressor effect of angiotensin I. These data indicate that noninvasive blood pressure monitoring at the finger represents a useful tool to study short-lasting blood pressure changes produced by vasoactive agents in man.
引用
收藏
页码:711 / 714
页数:4
相关论文
共 8 条
[1]   COMPLICATIONS OF PERCUTANEOUS RADIAL-ARTERY CANNULATION - OBJECTIVE PROSPECTIVE STUDY IN MAN [J].
BEDFORD, RF ;
WOLLMAN, H .
ANESTHESIOLOGY, 1973, 38 (03) :228-236
[2]  
BOEHMER RD, 1987, J CLIN MONITOR, V3, P282
[3]  
BURGENER E, 1988, J CARDIOVASC PHARM, V12, P126
[4]   HAZARDS OF RADIAL-ARTERY CATHETERIZATION [J].
DOWNS, JB ;
RACKSTEIN, AD ;
KLEIN, EF ;
HAWKINS, IF .
ANESTHESIOLOGY, 1973, 38 (03) :283-286
[5]  
FERGUSON RK, 1977, LANCET, V1, P775
[6]   EVALUATION OF THE PENAZ SERVO-PLETHYSMO-MANOMETER FOR THE CONTINUOUS, NON-INVASIVE MEASUREMENT OF FINGER BLOOD-PRESSURE [J].
MOLHOEK, GP ;
WESSELING, KH ;
SETTELS, JJM ;
VANVOLLENHOVEN, E ;
WEEDA, HWH ;
DEWIT, B ;
ARNTZENIUS, AC .
BASIC RESEARCH IN CARDIOLOGY, 1984, 79 (05) :598-609
[7]   COMPARISON OF FINGER AND INTRA-ARTERIAL BLOOD-PRESSURE MONITORING AT REST AND DURING LABORATORY TESTING [J].
PARATI, G ;
CASADEI, R ;
GROPPELLI, A ;
DIRIENZO, M ;
MANCIA, G .
HYPERTENSION, 1989, 13 (06) :647-655
[8]  
Penaz J., 1973, 10TH INT C MED BIOL, P104