EXTENDED EVALUATION OF THE TITANIUM GREENFIELD VENA-CAVAL FILTER

被引:59
作者
GREENFIELD, LJ
PROCTOR, MC
CHO, KJ
CUTLER, BS
FERRIS, EJ
MCFARLAND, D
SOBEL, M
TISNADO, J
机构
[1] UNIV MICHIGAN, DEPT SURG, ANN ARBOR, MI 48109 USA
[2] UNIV MICHIGAN, DEPT RADIOL, ANN ARBOR, MI 48109 USA
[3] UNIV MASSACHUSETTS, DEPT SURG, WORCESTER, MA 01605 USA
[4] UNIV ARKANSAS, DEPT RADIOL, LITTLE ROCK, AR 72204 USA
[5] VIRGINIA COMMONWEALTH UNIV MED COLL VIRGINIA, DEPT SURG, RICHMOND, VA USA
[6] VIRGINIA COMMONWEALTH UNIV MED COLL VIRGINIA, DEPT RADIOL, RICHMOND, VA USA
关键词
D O I
10.1016/0741-5214(94)90146-5
中图分类号
R61 [外科手术学];
学科分类号
摘要
Purpose: The purpose of this study was to evaluate the long-term safety and efficacy of the titanium Greenfield filter-modified hook for prevention of pulmonary embolism. Methods: We conducted a prospective study in 173 patients from four institutions who underwent clinical examination, abdominal radiography, and duplex ultrasound examinations of the vena cava and lower extremities. If indicated by protocol or clinical presentation, computed tomography scans, pulmonary angiograms, or venacavograms were obtained. Results: The most common procedural event was filter limb asymmetry (10%), which had no clinical significance. A variety of other minor procedural events occurred in another 10% of cases. Early follow-up (< 6 months) was completed in 149 patients, and long-term evaluation was completed in 113 (> 12 months). Deaths in 24 patients were from nonembolic causes in all but one. There were four suspected or confirmed recurrent pulmonary emboli, for an incidence of 3,5% (four of 113), with one death (0.9%). Pour patients had inferior vena cava occlusion at early follow-up and at long-term evaluation, only one remained occluded (1%). Insertion site venous thrombosis was seen in only two patients (2%). Conclusion: The titanium Greenfield filter provides protection comparable to the standard stainless steel Greenfield filter after 1 pear with a low incidence of recurrent pulmonary embolism (3.5%) and a high caval patency rate (99%).
引用
收藏
页码:458 / 465
页数:8
相关论文
共 8 条
[1]  
GREENFIELD L, IN PRESS CARDIOVASC
[2]  
GREENFIELD LJ, 1992, ARCH SURG-CHICAGO, V127, P969
[3]   EXPERIMENTAL EMBOLIC CAPTURE BY ASYMMETRIC GREENFIELD FILTERS [J].
GREENFIELD, LJ ;
PROCTOR, MC .
JOURNAL OF VASCULAR SURGERY, 1992, 16 (03) :436-444
[4]  
GREENFIELD LJ, 1988, SURGERY, V104, P706
[5]   EVOLUTION OF HOOK DESIGN FOR FIXATION OF THE TITANIUM GREENFIELD FILTER [J].
GREENFIELD, LJ ;
CHO, KJ ;
TAUSCHER, JR .
JOURNAL OF VASCULAR SURGERY, 1990, 12 (03) :345-353
[6]  
GREENFIELD LJ, 1981, ARCH SURG-CHICAGO, V116, P1451
[7]  
GREENFIELD LJ, 1991, J VASC SURG, V14, P253
[8]   GREENFIELD CAVAL FILTERS - LONG-TERM RADIOGRAPHIC FOLLOW-UP-STUDY [J].
MESSMER, JM ;
GREENFIELD, LJ .
RADIOLOGY, 1985, 156 (03) :613-618