CORRECTION OF MINERAL ACID INTERFERENCES IN INDUCTIVELY COUPLED PLASMA OPTICAL-EMISSION SPECTROMETRY ON COPPER AND MANGANESE USING INTERNAL STANDARDIZATION

被引:22
作者
GARDEN, LM
MARSHALL, J
LITTLEJOHN, D
机构
[1] ICI PLC,WILTON MAT RES CTR,MIDDLESBROUGH TS6 8JE,CLEVELAND,ENGLAND
[2] UNIV STRATHCLYDE,DEPT PURE & APPL CHEM,GLASGOW G1 1XL,SCOTLAND
关键词
INDUCTIVELY COUPLED PLASMA OPTICAL EMISSION SPECTROMETRY; MYERS-TRACY SIGNAL COMPENSATION; INTERNAL STANDARDIZATION; ACID INTERFERENCE EFFECT;
D O I
10.1039/ja9910600159
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
A study has been made of mineral acid matrix interferences in trace element determinations by inductively coupled plasma optical emission spectrometry. It has been shown that signal suppressions of up to 40% can be observed under normal operating conditions. The interferences can largely be compensated for by using an internal standardization procedure. Precision is also improved several fold when scandium is used as a real-time internal standard. This is attributed to the reduction of noise by simultaneous measurement of emission signals.
引用
收藏
页码:159 / 163
页数:5
相关论文
共 27 条
[1]   DETERMINATION OF COPPER IN BLOOD-SERUM BY DIRECT-CURRENT PLASMA AND INDUCTIVELY COUPLED PLASMA ATOMIC EMISSION-SPECTROMETRY [J].
BAMIRO, FO ;
LITTLEJOHN, D ;
MARSHALL, J .
JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 1988, 3 (01) :279-284
[2]   THEORETICAL PRINCIPLES OF INTERNAL STANDARDIZATION IN ANALYTICAL EMISSION SPECTROSCOPY [J].
BARNETT, WB ;
FASSEL, VA ;
KNISELEY, RN .
SPECTROCHIMICA ACTA PART B-ATOMIC SPECTROSCOPY, 1968, B 23 (10) :643-&
[3]   CORRELATION STUDY OF INTERNAL STANDARDIZATION IN INDUCTIVELY COUPLED PLASMA ATOMIC EMISSION-SPECTROMETRY [J].
BELCHAMBER, RM ;
HORLICK, G .
SPECTROCHIMICA ACTA PART B-ATOMIC SPECTROSCOPY, 1982, 37 (12) :1037-1046
[4]   INTERFERENCE FROM EASILY IONIZABLE ELEMENT MATRICES IN INDUCTIVELY COUPLED PLASMA EMISSION-SPECTROMETRY - A SPATIAL STUDY [J].
BLADES, MW ;
HORLICK, G .
SPECTROCHIMICA ACTA PART B-ATOMIC SPECTROSCOPY, 1981, 36 (09) :881-900
[6]   MODIFICATION AND OPTIMIZATION OF A 50 MHZ INDUCTIVELY COUPLED ARGON PLASMA WITH SPECIAL REFERENCE TO ANALYSES USING ORGANIC-SOLVENTS [J].
BOUMANS, PWJM ;
LUXSTEINER, MC .
SPECTROCHIMICA ACTA PART B-ATOMIC SPECTROSCOPY, 1982, 37 (02) :97-126
[7]   INDUCTIVELY COUPLED PLASMA-ATOMIC EMISSION SPECTROMETRY - ANALYSIS OF BIOLOGICAL-MATERIALS AND SOILS FOR MAJOR, TRACE, AND ULTRA-TRACE ELEMENTS [J].
DAHLQUIST, RL ;
KNOLL, JW .
APPLIED SPECTROSCOPY, 1978, 32 (01) :1-29
[8]   POSSIBILITIES FOR CORRECTING THE INFLUENCE OF MINERAL ACIDS AND OF DRIFT IN ICP EMISSION SPECTRAL-ANALYSIS [J].
DELIJSKA, A ;
VOUCHKOV, M .
FRESENIUS ZEITSCHRIFT FUR ANALYTISCHE CHEMIE, 1985, 321 (05) :448-452
[9]   INFLUENCE OF SOLUTION UPTAKE RATE ON SIGNALS AND INTERFERENCES IN INDUCTIVELY COUPLED PLASMA OPTICAL-EMISSION SPECTROMETRY [J].
FARINO, J ;
MILLER, JR ;
SMITH, DD ;
BROWNER, RF .
ANALYTICAL CHEMISTRY, 1987, 59 (18) :2303-2309
[10]   NEBULIZATION EFFECTS WITH ACID SOLUTIONS IN ICP SPECTROMETRY [J].
GREENFIELD, S ;
MCGEACHIN, HM ;
SMITH, PB .
ANALYTICA CHIMICA ACTA, 1976, 84 (01) :67-78