BIENZYME REACTOR FOR ELECTROCHEMICAL DETECTION OF LOW CONCENTRATIONS OF URIC-ACID AND GLUCOSE

被引:41
作者
ELEKES, O
MOSCONE, D
VENEMA, K
KORF, J
机构
[1] UNIV GRONINGEN,DEPT BIOL PSYCHIAT,9700 RB GRONINGEN,NETHERLANDS
[2] GEDEON RICHTER CHEM WORKS LTD,CHEM WORKS,PHARMACOL RES CTR,H-1475 BUDAPEST 10,HUNGARY
[3] UNIV ROMA TOR VERGATA,DIPARTIMENTO SCI & TECNOL CHIM,I-00133 ROME,ITALY
关键词
URIC ACID; GLUCOSE; ENZYME REACTOR; AMPEROMETRY; MEDIATOR; ASCORBIC ACID;
D O I
10.1016/0009-8981(95)06110-Y
中图分类号
R446 [实验室诊断]; R-33 [实验医学、医学实验];
学科分类号
1001 ;
摘要
An enzyme-based flow-injection amperometric analysis system (FIA) for monitoring of uric acid and glucose is described, The oxidase and peroxidase enzymes are physically co-immobilised in a sandwich-type reactor and ferrocene serves as a mediator. The assays are based on the measurement of a reduction current resulting from the enzymatic reactions, at a glassy carbon electrode held at 0.00 mV (vs, Ag/AgCl). The high selectivity (ascorbic acid did not interfere) is coupled to high sensitivity (a detection limit of 30 and 60 nmol/l for uric acid and glucose, respectively; signal/noise = 3) and good stability (the enzymes remained active for more than 6 weeks at 30 degrees C), The usefulness of the assay in clinical chemistry is illustrated by the measurement of human serum uric acid and glucose concentration. The results obtained were in fairly good agreement with those obtained using conventional hospital laboratory methods.
引用
收藏
页码:153 / 165
页数:13
相关论文
共 22 条
[1]  
BOUTELLE MG, 1992, ANAL CHEM, V64, P1792
[2]  
DEBOER J, 1994, J LAB CLIN MED, V124, P210
[3]  
DEBOER J, 1992, LANCET, V340, P547
[4]  
DUNCAN PH, 1982, CLIN CHEM, V28, P284
[5]   AN ENZYME-REACTOR FOR ELECTROCHEMICAL MONITORING OF CHOLINE AND ACETYLCHOLINE - APPLICATIONS IN HIGH-PERFORMANCE LIQUID-CHROMATOGRAPHY, BRAIN-TISSUE, MICRODIALYSIS AND CEREBROSPINAL-FLUID [J].
FLENTGE, F ;
VENEMA, K ;
KOCH, T ;
KORF, J .
ANALYTICAL BIOCHEMISTRY, 1992, 204 (02) :305-310
[6]   A METHOD FOR ESTIMATION OF HYDROGEN-PEROXIDE BASED ON MEDIATED ELECTRON-TRANSFER REACTIONS OF PEROXIDASES AT ELECTRODES [J].
FREW, JE ;
HARMER, MA ;
HILL, HAO ;
LIBOR, SI .
JOURNAL OF ELECTROANALYTICAL CHEMISTRY, 1986, 201 (01) :1-10
[7]   NOVEL, REAGENTLESS, AMPEROMETRIC BIOSENSOR FOR URIC-ACID BASED ON A CHEMICALLY-MODIFIED SCREEN-PRINTED CARBON ELECTRODE COATED WITH CELLULOSE-ACETATE AND URICASE [J].
GILMARTIN, MAT ;
HART, JP .
ANALYST, 1994, 119 (05) :833-840
[8]   CHEMICAL SENSORS [J].
JANATA, J ;
JOSOWICZ, M ;
DEVANEY, DM .
ANALYTICAL CHEMISTRY, 1994, 66 (12) :R207-R228
[9]   ACTION OF BIOLOGICALLY-RELEVANT OXIDIZING SPECIES UPON URIC-ACID - IDENTIFICATION OF URIC-ACID OXIDATION-PRODUCTS [J].
KAUR, H ;
HALLIWELL, B .
CHEMICO-BIOLOGICAL INTERACTIONS, 1990, 73 (2-3) :235-247
[10]   A SENSITIVE DETERMINATION OF URIC-ACID IN SERUM USING URICASE/CATALASE/FORMALDEHYDE DEHYDROGENASE COUPLED WITH FORMATE DEHYDROGENASE [J].
KAYAMORI, Y ;
KATAYAMA, Y .
CLINICAL BIOCHEMISTRY, 1994, 27 (02) :93-97