ENZYMATIC DETERMINATION OF L-LYSINE BY FLOW-INJECTION TECHNIQUES

被引:33
作者
POHLMANN, A
STAMM, WW
KUSAKABE, H
KULA, MR
机构
[1] UNIV DUSSELDORF,INST ENZYME TECHNOL,POB 2050,W-5170 JULICH,GERMANY
[2] YAMASA SHOYU CORP LTD,CHOSHI,CHIBA,JAPAN
关键词
Amino acids; Enzyme reactor; Lysine;
D O I
10.1016/S0003-2670(00)82091-7
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
An enzymatic assay that is highly selective for l-lysine, based on flow-injection techniques combined with spectrophotometric detection, is presented. l-Lysine-α-oxidase (E.C. 1.4.3.14) from Trichoderma viride and horseradish peroxidase were used in a coupled enzyme assay. Peroxide produced in the first reaction was converted by peroxidase with phenol and 4-aminoantipyrine to a quinoneimine dye detectable at 500 nm. An analytical enzyme reactor filled with coimmobilized enzymes was incorporated in the flow-injection system. The assay has a measuring frequency of 30 samples h-1 and a response time of less than 2 min. To adapt the assay to high concentrations of l-Lysine and to minimize interferences, the injected sample volume was reduced to 2 μ-l, resulting in a linearity range of 1-16 mM l-lysine with a sensitivity of 6-7 mV 1 mmol-1, a limit of detection (3σ) of 1 mM and a reproducibility of 0.5% (repetitive injection of a 10 mM l-lysine sample). The enzyme cartridge is stable for several months and thousands of measurements. © 1990.
引用
收藏
页码:329 / 335
页数:7
相关论文
共 45 条
[21]  
MERCK E, 1989, ASSAY DESCRIPTION PE
[22]  
MINODA Y, 1986, PROGR IND MICROBIOLO, V24, P51
[23]  
MOLLER J, 1988, ANAL TASCHENBUCH, V7, P199
[24]  
MOWERY RA, 1985, ISA T, V24, P1
[25]   ENZYMIC DETERMINATION OF L-LYSINE IN BIOLOGICAL-MATERIALS [J].
NAKATANI, Y ;
HIGASHINO, K ;
FUJIOKA, M .
ANALYTICAL BIOCHEMISTRY, 1972, 49 (01) :225-+
[26]   ONLINE FLOW-INJECTION MONITORING OF THE ENZYME INDUCTOR L-PHENYLALANINE IN THE CONTINUOUS CULTIVATION OF RHODOCOCCUS SP-M4 [J].
NALBACH, U ;
SCHIEMENZ, H ;
STAMM, WW ;
HUMMEL, W ;
KULA, MR .
ANALYTICA CHIMICA ACTA, 1988, 213 (1-2) :55-60
[27]  
NOPPER B, 1987, THESIS FACHHOCHSCHUL
[28]  
OLSSEN B, 1985, MIKROCHIM ACTA 2, P211
[29]   OPTIMIZATION OF PEROXIDASE IMMOBILIZATION AND OF THE DESIGN OF PACKED-BED ENZYME REACTORS FOR FLOW-INJECTION ANALYSIS [J].
OLSSON, B ;
OGREN, L .
ANALYTICA CHIMICA ACTA, 1983, 145 (JAN) :87-99
[30]   MULTIDETECTION FLOW-INJECTION TECHNIQUES FOR MANIPULATION OF SENSITIVITY - AMPLIFICATION AND DILUTION METHODS [J].
RIOS, A ;
LAZARO, F ;
DECASTRO, MDL ;
VALCARCEL, M .
ANALYTICA CHIMICA ACTA, 1987, 199 :15-27