MATCHING PHOTON AND ELECTRON FIELDS IN THE TREATMENT OF HEAD AND NECK TUMORS

被引:17
作者
PAPIEZ, E
DUNSCOMBE, PB
MALAKER, K
机构
[1] MANITOBA CANC TREATMENT & RES FDN,DEPT MED PHYS,WINNIPEG R3E 0V9,MANITOBA,CANADA
[2] MANITOBA CANC TREATMENT & RES FDN,DEPT RADIAT ONCOL,WINNIPEG R3E 0V9,MANITOBA,CANADA
关键词
TREATMENT PLANNING; COMBINED PHOTON AND ELECTRON BEAMS; PENUMBRA GENERATOR;
D O I
10.1118/1.596908
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Radiotherapy of head and neck tumors frequently involves joining photon and electron fields. In such situations narrow penumbras combined with relatively small positioning errors can lead to significant "hot" and "cold" spots in the vicinity of the join-up. The objective of this work was to devise penumbra spreading techniques which lead to a relatively uniform dose distribution in the join-up region of these fields and which reduce the effect of positioning errors on dose uniformity. A stepped edge attenuator was used to obtain a wider penumbra for the 4-MV x-ray beam and a Lucite scatterer was used for the 10-MeV electron beam. The resulting composite beam profiles from these "modified" abutting photon and electron fields are provided and the effects of positioning errors on dose uniformity across the junction are illustrated. These profiles are compared with those resulting from "unmodified" adjacent electron and photon beams.
引用
收藏
页码:335 / 341
页数:7
相关论文
共 6 条
[1]   A BEAM-EDGE MODIFIER FOR ABUTTING ELECTRON FIELDS [J].
KALEND, AM ;
ZWICKER, RD ;
WU, A ;
STERNICK, ES .
MEDICAL PHYSICS, 1985, 12 (06) :793-798
[2]  
KURUP RG, 1988, AAPM C SAN ANTONIO
[3]   FILM DOSIMETRY OF CLINICAL ELECTRON-BEAMS [J].
VANBATTUM, LJ ;
HUIZENGA, H .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1990, 18 (01) :69-76
[4]  
WANG CC, 1983, RAD THERAPY HEAD NEC, P185
[5]  
1984, ICRU35 INT COMM RAD, P23
[6]  
1983, MED PHYS, V6, P741