摘要
Treatment planning is a process requiring the cooperation of a number of different staff groups. The possibility for error is well recognised and quality control procedures are necessary to ensure that the chances of errors in planning leading to incorrect treatments are as low as possible. An audit system is described which is based on the calculation of the dose delivered to the patient using the parameters set for treatment as input data. The calculated dose is compared with the prescribed dose and errors greater than a defined limit are flagged. During a period of 19 months during which this audit procedure has been operating a total of 14 errors in excess of 5% were discovered and corrected, this is approximately 0.5% of the total number of plans checked.
引用
收藏
页码:197 / 202
页数:6
相关论文
共 9 条
[1]   QUALITY ASSURANCE CONTROL IN THE EORTC COOPERATIVE GROUP OF RADIOTHERAPY .1. ASSESSMENT OF RADIOTHERAPY STAFF AND EQUIPMENT [J].
HORIOT, JC ;
JOHANSSON, KA ;
GONZALEZ, DG ;
VANDERSCHUEREN, E ;
VANDENBOGAERT, W ;
NOTTER, G .
RADIOTHERAPY AND ONCOLOGY, 1986, 6 (04) :275-284
[2]   A RADIOTHERAPY TREATMENT PLANNING CHECK PROGRAM [J].
MORREY, D ;
SMITH, CW .
BRITISH JOURNAL OF RADIOLOGY, 1985, 58 (691) :668-670
[3]  
1989, MED ELECTRICAL EQUIP
[4]  
1988, IPSM54 I PHYS SCI ME
[5]  
1978, ICRU29 INT COMM RAD
[6]  
1984, AAPM13 AM ASS PHYS M
[7]  
1985, PHYS MED BIOL, V30, P1169
[8]  
1983, PHYS MED BIOL, V28, P1097
[9]  
1988, QUALITY ASSURANCE RA