THE DOTTER RETRIEVER AND PIGTAIL CATHETER - EFFICACY IN EXTRACTION OF CHRONIC TRANSVENOUS PACEMAKER LEADS

被引:22
作者
ESPINOSA, RE [1 ]
HAYES, DL [1 ]
VLIETSTRA, RE [1 ]
OSBORN, MJ [1 ]
MCGOON, MD [1 ]
机构
[1] WATSON CLIN,LAKELAND,FL
来源
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY | 1993年 / 16卷 / 12期
关键词
DOTTER; EXTRACTION; INFECTION; PACEMAKER; PIGTAIL; TRANSVENOUS ELECTRODE;
D O I
10.1111/j.1540-8159.1993.tb02347.x
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Several techniques exist for percutaneous extraction of chronic pacemaker leads. To establish the efficacy of the Dotter retriever and pigtail catheter, we reviewed the removal of 59 endocardial pacemaker leads in 42 patients (mean age 71 years). The mean duration of lead implantation was 44 months (range 1-169 months). Thirty-two leads were withdrawn with simple traction alone, and five leads were abandoned when traction failed. The remaining 22 leads were manipulated with a Dotter retriever or pigtail catheter, or both. Twelve leads were dislodged from the endocardium with simple traction (10) or with traction transmitted through an entwining pigtail catheter (2), but they could not be fully withdrawn. Eleven of these leads (92%) were then successfully extracted with the Dotter retriever. Seven of the remaining 10 leads were successfully dislodged and removed by the Dotter retriever. Overall, 9 of 12 leads (75%) that could not be dislodged from the endocardium with simple traction were removed with a Dotter retriever or pigtail catheter, or both. Three patients in whom no catheter method worked required thoracotomy for removal of infected leads. No complications resulted from use of the Dotter retriever or pigtail catheter. We conclude that the Dotter retriever and pigtail catheter have moderate efficacy for dislodging chronic endocardial leads. Once mobilized, however, the leads can be withdrawn with great success with the Dotter retriever. Newer technology should not result in the abandonment of this proven technique.
引用
收藏
页码:2337 / 2342
页数:6
相关论文
共 22 条
  • [1] CHRONIC TRANSVENOUS PACEMAKER LEAD REMOVAL USING A UNIQUE, SEQUENTIAL TRANSVENOUS SYSTEM
    BRODELL, GK
    CASTLE, LW
    MALONEY, JD
    WILKOFF, BL
    [J]. AMERICAN JOURNAL OF CARDIOLOGY, 1990, 66 (12) : 964 - 966
  • [2] BRODMAN R, 1992, J THORAC CARDIOV SUR, V103, P649
  • [3] BYRD CL, 1991, J THORAC CARDIOV SUR, V101, P989
  • [4] TRANSLUMINAL EXTRACTION OF CATHETER AND GUIDE FRAGMENTS FROM HEART AND GREAT VESSELS - 29 COLLECTED CASES
    DOTTER, CT
    ROSCH, J
    BILBAO, MK
    [J]. AMERICAN JOURNAL OF ROENTGENOLOGY RADIUM THERAPY AND NUCLEAR MEDICINE, 1971, 111 (03): : 467 - +
  • [5] INTRAVASCULAR LEAD EXTRACTION USING LOCKING STYLETS, SHEATHS, AND OTHER TECHNIQUES
    FEARNOT, NE
    SMITH, HJ
    GOODE, LB
    BYRD, CL
    WILKOFF, BL
    SELLERS, TD
    [J]. PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1990, 13 (12): : 1864 - 1870
  • [6] PERCUTANEOUS REMOVAL OF VENTRICULAR PACEMAKER ELECTRODES USING A DORMIER BASKET
    FOSTER, CJ
    BROWNLEE, WC
    [J]. INTERNATIONAL JOURNAL OF CARDIOLOGY, 1988, 21 (02) : 127 - 134
  • [7] FURMAN S, 1987, J THORAC CARDIOV SUR, V94, P770
  • [8] FURMAN S, 1989, PRACTICE CARDIAC PAC, P633
  • [9] FORCEPS EXTRACTION OF PERMANENT PACING LEADS
    KRATZ, JM
    LEMAN, R
    GILLETTE, PC
    [J]. ANNALS OF THORACIC SURGERY, 1990, 49 (04) : 676 - 677
  • [10] REMOVAL OF AN IATROGENIC FOREIGN BODY FROM AORTA BY MEANS OF A URETERIC STONE CATCHER
    LASSERS, BW
    PICKERING, D
    [J]. AMERICAN HEART JOURNAL, 1967, 73 (03) : 375 - +