FEMRI GEL;
RADIATION THERAPY;
3D TREATMENT PLANNING;
DOSE MAPPING;
MRI;
D O I:
10.1118/1.597612
中图分类号:
R8 [特种医学];
R445 [影像诊断学];
学科分类号:
1002 ;
100207 ;
1009 ;
摘要:
A detailed methodology has been developed to verify the three-dimensional (3D) radiation dose mapping of conformal therapy radiation treatment planning by FeMRI dosimetry. A phantom that consisted of a human skull filled with a 1-mM solution of ferrous sulfate and 0.1-N sulfuric acid gelled with 7.5% (by weight) gelatin was employed. With a spherical target volume in the head phantom, five noncoplanar conformal beams were designed through the use of a 3D treatment planning system developed in-house. The phantom was irradiated with a 6-MV linear accelerator to a total dose of 25 Gy delivered to the periphery of the target volume. The phantom and a set of calibration vials were scanned simultaneously in a GE 1.5T MR imager with six different multiscan inversion-recovery pulse sequences. The values of T-1 were evaluated on a pixel-by-pixel basis through the use of custom-built software on a UNIX workstation and were converted to dose using calibration data. Comparisons of dose distributions between those measured by FeMRI and those calculated by 3D treatment planning show good agreement.