TRIPLE PHOTON-ABSORPTIOMETRY CANNOT CORRECT FOR FAT INHOMOGENEITIES IN LUMBAR SPINE BONE-MINERAL MEASUREMENTS

被引:14
作者
FARRELL, TJ [1 ]
WEBBER, CE [1 ]
机构
[1] CHEDOKE MCMASTER HOSP,DEPT NUCL MED,1200 MAIN ST W,HAMILTON L8N 3Z5,ONTARIO,CANADA
来源
CLINICAL PHYSICS AND PHYSIOLOGICAL MEASUREMENT | 1990年 / 11卷 / 01期
关键词
D O I
10.1088/0143-0815/11/1/007
中图分类号
Q6 [生物物理学];
学科分类号
071011 ;
摘要
In vivo fat inhomogeneities can cause significant errors in measurements of lumbar spine mineral mass using dual photon absorptiometry. Extension to a triple photon techniques to correct for the presence of fat is not possible because of the unacceptable increase of about a factor of 3000 required in the radiation dose to yield sufficient transmitted photons for adequate precision.
引用
收藏
页码:77 / 84
页数:8
相关论文
共 17 条
  • [1] ENERGY-SELECTIVE RECONSTRUCTIONS IN X-RAY COMPUTERIZED TOMOGRAPHY
    ALVAREZ, RE
    MACOVSKI, A
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 1976, 21 (05) : 733 - 744
  • [2] [Anonymous], 1969, DATA REDUCTION ERROR
  • [3] ARVIN DE, 1978, INVEST RADIOL, V13, P217
  • [4] BILLINGSLEY FC, 1975, PICTURE PROCESSING D
  • [5] THE ERROR DUE TO FAT INHOMOGENEITY IN LUMBAR SPINE BONE-MINERAL MEASUREMENTS
    FARRELL, TJ
    WEBBER, CE
    [J]. CLINICAL PHYSICS AND PHYSIOLOGICAL MEASUREMENT, 1989, 10 (01): : 57 - 64
  • [6] METHODOLOGIES FOR THE MEASUREMENT OF BONE-DENSITY AND THEIR PRECISION AND ACCURACY
    GOODWIN, PN
    [J]. SEMINARS IN NUCLEAR MEDICINE, 1987, 17 (04) : 293 - 304
  • [7] TISSUE-ANALYSIS BY DUAL-ENERGY COMPUTED-TOMOGRAPHY
    HAWKES, DJ
    JACKSON, DF
    PARKER, RP
    [J]. BRITISH JOURNAL OF RADIOLOGY, 1986, 59 (702) : 537 - 542
  • [8] PHOTON MASS ATTENUATION AND ENERGY-ABSORPTION COEFFICIENTS FROM 1 KEV TO 20 MEV
    HUBBELL, JH
    [J]. INTERNATIONAL JOURNAL OF APPLIED RADIATION AND ISOTOPES, 1982, 33 (11): : 1269 - 1290
  • [9] JACKSON DF, 1983, PHYS MED BIOL, V28, P288
  • [10] BONE-MINERAL MEASUREMENT WITH A CONTINUOUS ROENTGEN RAY SPECTRUM AND A GERMANIUM DETECTOR
    JONSON, R
    ROOS, B
    HANSSON, T
    [J]. ACTA RADIOLOGICA-DIAGNOSIS, 1986, 27 (01): : 105 - 109