Use of poly(ester-sulfonic acid) film-coated electrodes as amperometric detectors in flow injection analysis

被引:1
作者
Gennett, T
Rusin, WR
Gier, LM
机构
[1] Chemistry Department, Rochester Institute of Technology, Rochester, NY 14623
关键词
chemically modified electrodes; amperometric detectors; poly(ester-sulfonic acid); flow injection analysis;
D O I
10.1016/0039-9140(95)01689-9
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
Within the past several years significant advances have been made towards the development and incorporation of chemically modified electrodes as selective detectors for high performance liquid chromatography and flow injection analysis, In many cases the chemically modified electrode systems closely approach the ''ideal'' detector specifications of chemical and mechanical stability along with a significant linear response region, This paper will discuss the characterization and incorporation of ionomeric poly(ester-sulfonic acid) coated electrodes as nonaqueous electrochemical detectors. The orientation of the electrodes in the detector system as well as the increased sensitivity levels to 10(-10) g ml(-1) for cationic species and 10(-9) g ml(-1) for neutral species will be presented. Also the applicability of the ionomer coated electrodes as nonelectrolyte detectors achieved a reproducible response with detection limits to 10(-6) g ml(-1). Overall this system performed as well as, or better than, more specialized and expensive thin layer electrochemical detectors.
引用
收藏
页码:2073 / 2078
页数:6
相关论文
共 17 条
[1]   AN EVALUATION OF ULTRATHIN RING AND BAND MICROELECTRODES AS AMPEROMETRIC SENSORS IN ELECTROCHEMICAL FLOW CELLS [J].
BIXLER, JW ;
FIFIELD, M ;
POLER, JC ;
BOND, AM ;
THORMANN, W .
ELECTROANALYSIS, 1989, 1 (01) :23-33
[2]   AMPEROMETRIC DETECTION OF PICOMOLE SAMPLES IN A MICRODISK ELECTROCHEMICAL FLOW-JET CELL WITH DILUTE SUPPORTING ELECTROLYTE [J].
BIXLER, JW ;
BOND, AM .
ANALYTICAL CHEMISTRY, 1986, 58 (13) :2859-2863
[3]   SENSORS FROM POLYMER MODIFIED ELECTRODES [J].
ESPENSCHEID, MW ;
GHATAKROY, AR ;
MOORE, RB ;
PENNER, RM ;
SZENTIRMAY, MN ;
MARTIN, CR .
JOURNAL OF THE CHEMICAL SOCIETY-FARADAY TRANSACTIONS I, 1986, 82 :1051-1070
[4]   PERMEABILITY OF GLUCOSE AND OTHER NEUTRAL SPECIES THROUGH RECAST PERFLUOROSULFONATED IONOMER FILMS [J].
FAN, ZH ;
HARRISON, DJ .
ANALYTICAL CHEMISTRY, 1992, 64 (11) :1304-1311
[5]   VOLTAMMETRIC DETERMINATION OF THE ION-EXCHANGE BEHAVIOR OF POLY(ESTER SULFONIC-ACID) ANIONOMERS IN ACETONITRILE [J].
GENNETT, T ;
PURDY, WC .
ANALYTICAL CHEMISTRY, 1990, 62 (19) :2155-2158
[6]   CHARACTERIZATION OF PERFLUOROSULFONIC ACID POLYMER COATED ENZYME ELECTRODES AND A MINIATURIZED INTEGRATED POTENTIOSTAT FOR GLUCOSE ANALYSIS IN WHOLE-BLOOD [J].
HARRISON, DJ ;
TURNER, RFB ;
BALTES, HP .
ANALYTICAL CHEMISTRY, 1988, 60 (19) :2002-2007
[7]   ELECTROCHEMICAL FABRICATION OF AMPEROMETRIC MICROENZYME SENSOR [J].
IKARIYAMA, Y ;
YAMAUCHI, S ;
YUKIASHI, T ;
USHIODA, H .
JOURNAL OF THE ELECTROCHEMICAL SOCIETY, 1989, 136 (03) :702-706
[8]   VOLTAMMETRY WITHOUT SUPPORTING ELECTROLYTE USING A PLATINUM WORKING ELECTRODE SUPPORTED ON AN ION-EXCHANGE MEMBRANE [J].
KAARET, TW ;
EVANS, DH .
ANALYTICAL CHEMISTRY, 1988, 60 (07) :657-662
[9]  
LANGMIRE ML, 1990, ANAL CHEM, V62, P747
[10]  
SZENTIMAY MN, 1984, ANAL CHEM, V56, P1989