AN IMPROVED DESIGN OF VOLTMETER FOR SEMI-PARALLEL DATA ACQUISITION

被引:10
作者
TRILLAUD, C
JOSSINET, J
机构
[1] INSERM U281, Lyon Cedex, 69424
来源
CLINICAL PHYSICS AND PHYSIOLOGICAL MEASUREMENT | 1992年 / 13卷
关键词
D O I
10.1088/0143-0815/13/A/001
中图分类号
Q6 [生物物理学];
学科分类号
071011 ;
摘要
The design and implementation of high performance differential voltmeters for semi-parallel data acquisition are described. The general requirements and specific conditions encountered in electrical impedance tomography (EIT) for accurate measurements are analysed. The major parameters are common-mode rejection and bandwidth. A specific implementation of the voltmeters, with separate DC supplies and independent signal references, is described. This arrangement, in which each voltmeter follows the input signal, automatically cancels any common-mode voltage present at the input. The signal is fed to the remainder of the instrumentation through a transformer. The use of a reduced number of components contributes to the minimisation of the inter-channel variations. Furthermore, the geometrical distribution of the voltmeters around the object minimises the length of electrode wires, also reducing the input capacitance. The number of modular voltmeters and DC/DC converters is 32. The common-mode rejection of these voltmeters is greater than 72 dB in the frequency range 3.6-560 kHz. In conclusion, the proposed solution ensures a minimisation of common-mode errors and enables the use of a 250 kHz frequency.
引用
收藏
页码:5 / 10
页数:6
相关论文
共 12 条
  • [1] APPLIED POTENTIAL TOMOGRAPHY
    BARBER, DC
    BROWN, BH
    [J]. JOURNAL OF PHYSICS E-SCIENTIFIC INSTRUMENTS, 1984, 17 (09): : 723 - 733
  • [2] A DUAL-FREQUENCY ELECTRICAL-IMPEDANCE TOMOGRAPHY SYSTEM
    GRIFFITHS, H
    ZHANG, Z
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 1989, 34 (10) : 1465 - 1476
  • [3] HENDERSON RP, 1985, IEEE T BIOMED ENG, V32, P177
  • [4] JOSSINET J, 1990, P 3 EUR COMM WORKSH, P144
  • [5] Aspects of instrumentation design for impedance imaging
    Murphy, D.
    Rolfe, P.
    [J]. Clinical Physics and Physiological Measurement, 1988, 9 (SUPPL. A): : 5 - 14
  • [6] AN ELECTRIC-CURRENT TOMOGRAPH
    NEWELL, JC
    GISSER, DG
    ISAACSON, D
    [J]. IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 1988, 35 (10) : 828 - 833
  • [8] ROSELL J, 1989, 11TH P INT ANN C IEE, V11, P459
  • [9] SARANUMMI NK, 1976, 11TH P INT C BIOM EN, V33, P460
  • [10] TRILLAUD C, 1991, 1ST P EUR C BIOM ENG, V36