CHEST RADIOGRAPHY - COMPARISON OF HIGH-RESOLUTION DIGITAL DISPLAYS WITH CONVENTIONAL AND DIGITAL FILM

被引:73
作者
COX, GG
COOK, LT
MCMILLAN, JH
ROSENTHAL, SJ
DWYER, SJ
机构
[1] Department of Diagnostic Radiology, University of Kansas Medical Center, Kansas City, KS 66103
关键词
Diagnostic radiology; observer performance; Radiography; comparative studies; computer-assisted; digital; Thorax; radiography; 60.11; 60.1215;
D O I
10.1148/radiology.176.3.2389035
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
This study was performed to compare the performance of observers using three display formats for chest radiography. The display formats were conventional radiographs, digitized radiographs (2,048 × 2,048 × 12 bits) printed on laser film, and digitized radiographs (2,048 × 2,048 × 12 bits) displayed on a high-resolution (2,560 × 2,048 × 12-bit) gray-scale display. The test set for the study consisted of 163 cases. Sixty-four of the cases were normal, whereas the 99 remaining cases demonstrated one or more common radiographic abnormalities. Nine abnormalities were selected for analysis: costophrenic angle blunting, interstitial disease, atelectasis, pneumothorax, parenchymal mass, consolidation, obstructive disease, hilar/mediastinal mass, and apical scarring. Six experienced general radiologists participated in the evaluation. Receiver operating characteristic curves were generated for each abnormality and display format. The results indicate that, while the three display formats are equivalent for the detection of some abnormalities, detectable differences in observer performance may be seen even at 2,048 × 2,048 × 12 bits for the detection of obstructive disease, pneumothorax, interstitial disease, and parenchymal masses.
引用
收藏
页码:771 / 776
页数:6
相关论文
共 15 条
[1]   DIGITAL AND CONVENTIONAL CHEST IMAGING - A MODIFIED ROC STUDY OF OBSERVER PERFORMANCE USING SIMULATED NODULES [J].
CHAKRABORTY, DP ;
BREATNACH, ES ;
YESTER, MV ;
SOTO, B ;
BARNES, GT ;
FRASER, RG .
RADIOLOGY, 1986, 158 (01) :35-39
[2]   THE EFFECT OF VARYING SPATIAL-RESOLUTION ON THE DETECTABILITY OF DIFFUSE PULMONARY NODULES - ASSESSMENT WITH DIGITIZED CONVENTIONAL RADIOGRAPHS [J].
FOLEY, WD ;
WILSON, CR ;
KEYES, GS ;
DIBIANCA, FA ;
SCANLON, GT ;
SCHLEUTER, D ;
LAWSON, TL .
RADIOLOGY, 1981, 141 (01) :25-31
[3]  
FRANK MS, 1989, RADIOLOGY P, V173, P401
[4]   DIGITAL AND CONVENTIONAL CHEST IMAGES - OBSERVER PERFORMANCE WITH FILM DIGITAL RADIOGRAPHY SYSTEM [J].
GOODMAN, LR ;
FOLEY, WD ;
WILSON, CR ;
RIMM, AA ;
LAWSON, TL .
RADIOLOGY, 1986, 158 (01) :27-33
[5]  
HABICHT JP, 1980, SCIENCE, V207, P1414, DOI 10.1126/science.7361090
[6]  
HANLEY JA, 1989, CRIT REV DIAGN IMAG, V29, P307
[7]   COMPARISON OF 2048-LINE DIGITAL-DISPLAY FORMATS AND CONVENTIONAL RADIOGRAPHS - AN ROC STUDY [J].
HAYRAPETIAN, A ;
ABERLE, DR ;
HUANG, HK ;
FISKE, R ;
MORIOKA, C ;
VALENTINO, D ;
BOECHAT, MI .
AMERICAN JOURNAL OF ROENTGENOLOGY, 1989, 152 (05) :1113-1118
[8]   SPATIAL-RESOLUTION REQUIREMENTS FOR DIGITAL CHEST RADIOGRAPHS - AN ROC STUDY OF OBSERVER PERFORMANCE IN SELECTED CASES [J].
LAMS, PM ;
COCKLIN, ML .
RADIOLOGY, 1986, 158 (01) :11-19
[9]   DIGITAL RADIOGRAPHY OF SUBTLE PULMONARY ABNORMALITIES - AN ROC STUDY OF THE EFFECT OF PIXEL SIZE ON OBSERVER PERFORMANCE [J].
MACMAHON, H ;
VYBORNY, CJ ;
METZ, CE ;
DOI, K ;
SABETI, V ;
SOLOMON, SL .
RADIOLOGY, 1986, 158 (01) :21-26
[10]  
METZ CE, 1986, INVEST RADIOL, V21, P720, DOI 10.1097/00004424-198609000-00009