CINERADIOGRAPHY FOR DETERMINATION OF NORMAL AND ABNORMAL FUNCTION IN MECHANICAL HEART-VALVES

被引:23
作者
VOGEL, W [1 ]
STOLL, HP [1 ]
BAY, W [1 ]
FROHLIG, G [1 ]
SCHIEFFER, H [1 ]
机构
[1] UNIV HOMBURG,MED KLIN & POLIKLIN INNERE MED 3 KARDIOL,W-6650 HOMBURG,GERMANY
关键词
D O I
10.1016/0002-9149(93)90742-U
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
To determine the diagnostic value af cineradiography of mechanical heart valves, 112 cinefluoroscopic studies were performed in 76 patients with 95 valve prostheses (caged ball or disk valves, tilting disk and bileaflet valves). A patient group (n = 45) presenting with clinical or echocardiographic findings suggestive of valve-related complications was compared with a control group (n = 31) without such symptoms. Disk-opening angles (mean +/- SD) for Medtronic Hall aortic valves were found to be significantly smaller (62.8 +/- 11.1-degrees) in patients than in control subjects (73.9 +/- 1.6-degrees; p <0.05). Tissue ingrowth or thrombus formation, or both, demonstrated in 3 patients on subsequent reoperation, are considered as the main cause of incomplete or asymmetric disk opening. Opening and closing times did not differ significantly between patients and control subjects. Besides abnormal valve motion, structural defects such as strut fracture or leaflet escape could be rapidly detected by cineradiography if x-ray projections according to the particular valve design were used. Together with quantitative Doppler echocardiographic and clinical data, this method can help to give specific answers if the question is to either confirm or exclude imminent or acute valve malfunction. Thus, modern cineradiography is a highly valuable noninvasive diagnostic tool for both rapid management of emergency cases and routine follow-up of patients with mechanical heart valves.
引用
收藏
页码:225 / 232
页数:8
相关论文
共 25 条
  • [1] CINERADIOGRAPHIC FUNCTIONAL-EVALUATION OF THE BJORK-SHILEY MONOSTRUT PROSTHESIS - A FOLLOW-UP-STUDY OF 46 CONSECUTIVE PATIENTS
    ALBRECHTSSON, UG
    THULIN, LI
    OLIN, CL
    [J]. SCANDINAVIAN JOURNAL OF THORACIC AND CARDIOVASCULAR SURGERY, 1987, 21 (01): : 91 - 95
  • [2] BURCKHARDT D, 1983, ADV CARDIAC VALVES, P199
  • [3] INVIVO RADIOGRAPHIC APPEARANCE OF THE ST-JUDE VALVE PROSTHESIS
    CASTANEDAZUNIGA, W
    NICOLOFF, D
    JORGENSEN, C
    NATH, PH
    ZOLLIKOFER, C
    AMPLATZ, K
    [J]. RADIOLOGY, 1980, 134 (03) : 775 - 776
  • [4] NEW CINERADIOGRAPHIC METHODS FOR ASSESSMENT OF BJORK-SHILEY VALVE INTEGRITY
    COUSIN, AJ
    SOTO, B
    [J]. AMERICAN JOURNAL OF ROENTGENOLOGY, 1986, 147 (06) : 1139 - 1144
  • [5] FIBRINOLYTIC THERAPY OF ST-JUDE VALVE THROMBOSIS UNDER GUIDANCE OF DIGITAL CINEFLUOROSCOPY
    CZER, LSC
    WEISS, M
    BATEMAN, TM
    PFAFF, JM
    DEROBERTIS, M
    EIGLER, N
    VAS, R
    MATLOFF, JM
    GRAY, RJ
    [J]. JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 1985, 5 (05) : 1244 - 1249
  • [6] CINEFLUOROSCOPIC EVALUATION OF PERIPROSTHETIC CARDIAC-VALVE REGURGITATION
    GREEN, CE
    GLASSROYAL, M
    BREAM, PR
    SOTO, B
    ELLIOTT, LP
    [J]. AMERICAN JOURNAL OF ROENTGENOLOGY, 1988, 151 (03) : 455 - 459
  • [7] HARTZ RS, 1986, J THORAC CARDIOV SUR, V92, P684
  • [8] OUTLET STRUT FRACTURE OF THE BJORK-SHILEY 60-DEGREES CONVEXO-CONCAVE VALVE - CURRENT INFORMATION AND RECOMMENDATIONS FOR PATIENT-CARE
    HIRATZKA, LF
    KOUCHOUKOS, NT
    GRUNKEMEIER, GL
    MILLER, DC
    SCULLY, HE
    WECHSLER, AS
    [J]. JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 1988, 11 (05) : 1130 - 1137
  • [9] CINERADIOGRAPHIC EVALUATION OF THE CONVEXO-CONCAVE BJORK-SHILEY PROSTHETIC VALVE IN MITRAL POSITION
    ISHIMARU, S
    FURUAWKA, K
    TAKAHASHI, M
    [J]. SCANDINAVIAN JOURNAL OF THORACIC AND CARDIOVASCULAR SURGERY, 1983, 17 (03): : 211 - 215
  • [10] LINDBLOM D, 1986, J THORAC CARDIOV SUR, V92, P894