ACCURATE COMBINATION OF CT AND MR DATA OF THE HEAD - VALIDATION AND APPLICATIONS IN SURGICAL AND THERAPY PLANNING

被引:31
作者
HILL, DLG [1 ]
HAWKES, DJ [1 ]
HUSSAIN, Z [1 ]
GREEN, SEM [1 ]
RUFF, CF [1 ]
ROBINSON, GP [1 ]
机构
[1] UMDS, GUYS HOSP, DEPT NEUROL, LONDON, ENGLAND
关键词
MULTIMODALITY; IMAGE REGISTRATION; VISUALIZATION; MR; CT;
D O I
10.1016/0895-6111(93)90029-M
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
A method is presented for the accurate combination of magnetic resonance (MR) and computed tomographic (CT) images of the head. Our technique is based on user identified 3D landmarks followed by data combination and display as adjacent slices, a single fused slice representation, colour overlay and three-dimensional (3D) rendered scenes. Validation with a point phantom and computer simulation has established the relationship of registration accuracy with point location accuracy, the number of points used and their spatial distribution. The technique is in clinical use in the planning of skull base surgery, transferring MR images acquired without a stereotaxic frame to stereotaxic space, and staging and planning therapy of nasopharyngeal tumours.
引用
收藏
页码:357 / 363
页数:7
相关论文
共 15 条
[1]  
ADAMS L, 1990, NATO ADV SCI I F-COM, V60, P411
[2]   LEAST-SQUARES FITTING OF 2 3-D POINT SETS [J].
ARUN, KS ;
HUANG, TS ;
BLOSTEIN, SD .
IEEE TRANSACTIONS ON PATTERN ANALYSIS AND MACHINE INTELLIGENCE, 1987, 9 (05) :699-700
[3]   MRI-PET CORRELATION IN 3 DIMENSIONS USING A VOLUME-OF-INTEREST (VOI) ATLAS [J].
EVANS, AC ;
MARRETT, S ;
TORRESCORZO, J ;
KU, S ;
COLLINS, L .
JOURNAL OF CEREBRAL BLOOD FLOW AND METABOLISM, 1991, 11 (02) :A69-A78
[4]  
GUTHRIE B L, 1990, Stereotactic and Functional Neurosurgery, V54-55, P497
[5]   MULTIMODALITY IMAGE INTEGRATION FOR STEREOTAXIC SURGICAL PLANNING [J].
HENRI, CJ ;
COLLINS, DL ;
PETERS, TM .
MEDICAL PHYSICS, 1991, 18 (02) :167-177
[6]   REGISTRATION OF MR AND CT IMAGES FOR SKULL BASE SURGERY USING POINT-LIKE ANATOMICAL FEATURES [J].
HILL, DLG ;
HAWKES, DJ ;
CROSSMAN, JE ;
GLEESON, MJ ;
COX, TCS ;
BRACEY, EECML ;
STRONG, AJ ;
GRAVES, P .
BRITISH JOURNAL OF RADIOLOGY, 1991, 64 (767) :1030-1035
[7]  
LEMOINE D, 1991, INFORMATION PROCESSI
[8]  
NEELIN P, 1992, P SAT S 3D ADV IM PR, P73
[9]   ACCURATE 3-DIMENSIONAL REGISTRATION OF CT, PET, AND OR MR IMAGES OF THE BRAIN [J].
PELIZZARI, CA ;
CHEN, GTY ;
SPELBRING, DR ;
WEICHSELBAUM, RR ;
CHEN, CT .
JOURNAL OF COMPUTER ASSISTED TOMOGRAPHY, 1989, 13 (01) :20-26
[10]  
PIETRZYK U, 1990, J COMPUT ASSIST TOMO, V14, P51, DOI 10.1097/00004728-199001000-00009