A COMPARISON OF 2 DUAL-ENERGY X-RAY ABSORPTIOMETRY SYSTEMS FOR SPINAL BONE-MINERAL MEASUREMENT

被引:33
作者
LAI, KC [1 ]
GOODSITT, MM [1 ]
MURANO, R [1 ]
CHESNUT, CH [1 ]
机构
[1] UNIV WASHINGTON, OSTEOPOROSIS RES CTR,DEPT RADIOL,DIV NUCL MED, RC-70, SEATTLE, WA 98195 USA
关键词
DUAL-ENERGY X-RAY ABSORPTIOMETRY; BONE MINERAL CONTENT; BONE MINERAL DENSITY; LINEARITY; EXPOSURE;
D O I
10.1007/BF00296283
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Two dual-energy X-ray absorptiometry (DEXA) systems-the Hologic QDR-1000 and the Norland XR-26 bone densitometers-were evaluated in terms of precision, accuracy, linearity of response, X-ray exposure, and correlation of in vivo spinal measurements. In vitro precision and accuracy studies were performed using the Hologic anthropomorphic spine phantom; linearity of response was determined with increasing thicknesses of aluminum slabs and concentrations of Tums E-X in a constant-level water bath. Both systems were comparable in precision, achieving coefficients of variation (CVs) of less than 1% in bone mineral content (BMC, g), bone area (cm2), and bone mineral density (BMD, g/cm2). Both were accurate in their determination of BMC, bone area, and BMD with reference to the Hologic spine phantom. Both systems also showed good BMC and BMD linearity of response. Measured X-ray skin surface exposures for the Hologic and the Norland systems were 3.11 and 3.02 mR, respectively. In vivo spinal measurements (n = 65) on the systems were highly correlated (BMC: r = 0.993, SEE = 1.770 g; area: r = 0.984, SEE = 1.713 cm2; BMD: r = 0.990, SEE = 0.028 g/cm2). In conclusion, both systems are comparable in terms of precision, accuracy, linearity of response, and exposure efficiency.
引用
收藏
页码:203 / 208
页数:6
相关论文
共 23 条
[1]   ENERGY-SELECTIVE RECONSTRUCTIONS IN X-RAY COMPUTERIZED TOMOGRAPHY [J].
ALVAREZ, RE ;
MACOVSKI, A .
PHYSICS IN MEDICINE AND BIOLOGY, 1976, 21 (05) :733-744
[2]  
ARAI H, 1990, IMAGE TECHNOL INFORM, V22, P1
[3]   A METHOD FOR SELECTIVE TISSUE AND BONE VISUALIZATION USING DUAL ENERGY SCANNED PROJECTION RADIOGRAPHY [J].
BRODY, WR ;
BUTT, G ;
HALL, A ;
MACOVSKI, A .
MEDICAL PHYSICS, 1981, 8 (03) :353-357
[4]   MEASUREMENT OF BONE MINERAL IN VIVO - AN IMPROVED METHOD [J].
CAMERONJR. ;
SORENSON, J .
SCIENCE, 1963, 142 (358) :230-+
[5]   PRECISE MEASUREMENT OF VERTEBRAL MINERAL-CONTENT USING COMPUTED-TOMOGRAPHY [J].
CANN, CE ;
GENANT, HK .
JOURNAL OF COMPUTER ASSISTED TOMOGRAPHY, 1980, 4 (04) :493-500
[6]   BONE COMPRESSIVE STRENGTH - INFLUENCE OF DENSITY AND STRAIN RATE [J].
CARTER, DR ;
HAYES, WC .
SCIENCE, 1976, 194 (4270) :1174-1176
[7]   COMPARISON OF 4 DUAL ENERGY IMAGE DECOMPOSITION METHODS [J].
CHUANG, KS ;
HUANG, HK .
PHYSICS IN MEDICINE AND BIOLOGY, 1988, 33 (04) :455-466
[8]   X-RAY DUAL-PHOTON ABSORPTIOMETRY - A NEW METHOD FOR THE MEASUREMENT OF BONE-DENSITY [J].
CULLUM, ID ;
ELL, PJ ;
RYDER, JP .
BRITISH JOURNAL OF RADIOLOGY, 1989, 62 (739) :587-592
[9]   QUANTITATIVE BONE-MINERAL ANALYSIS USING DUAL ENERGY COMPUTED TOMOGRAPHY [J].
GENANT, HK ;
BOYD, D .
INVESTIGATIVE RADIOLOGY, 1977, 12 (06) :545-551
[10]   DUAL-ENERGY RADIOGRAPHIC ABSORPTIOMETRY OF THE LUMBAR SPINE - CLINICAL-EXPERIENCE WITH 2 DIFFERENT SYSTEMS [J].
GUNDRY, CR ;
MILLER, CW ;
RAMOS, E ;
MOSCONA, A ;
STEIN, JA ;
MAZESS, RB ;
SARTORIS, DJ ;
RESNICK, D .
RADIOLOGY, 1990, 174 (02) :539-541