EFFECTS OF ANALYTE VELOCITY MODULATION ON THE ELECTROOSMOTIC FLOW IN CAPILLARY ELECTROPHORESIS

被引:9
作者
DEMANA, T [1 ]
GUHATHAKURTA, U [1 ]
MORRIS, MD [1 ]
机构
[1] UNIV MICHIGAN,DEPT CHEM,ANN ARBOR,MI 48109
关键词
D O I
10.1021/ac00028a013
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
Modulation of the electroosmotic flow in capillary zone electrophoresis by modulation of the driving voltage gives rise to a flow profile that changes between laminar and flat profiles. The changing flow profile induces a radial movement of sample species to and from the capillary surface. The induced sample concentration gradient can be monitored by carefully probing the capillary surface. The resulting signal is a derivative of the normal-shaped peak. Derivative-shaped peaks can be observed with cations, but not with anions. Anions are unable to approach the double-layer region and therefore are unaffected by the modulation process.
引用
收藏
页码:390 / 394
页数:5